5
• Testing a Blood Sample
86
8
The test strip icon prompts you to insert a test strip.
Remove a test strip from its container and close the
container again with the stopper.
9
Hold the test strip so the lettering with the test
name is facing upward.
10
Slide the test strip into the test strip guide in the
direction indicated by the arrows.
Slide the test strip in as far as it will go. A beep indi-
cates that the meter has detected the test strip
(provided the beeper is enabled).
Test
09:15 am
04/19/2012
P-ID: PID01
Müller, David
Op:
Schulz
Par:
PT
Exposure to external influences (such as humidity) may
deteriorate the test strips and may lead to error mes-
sages. Therefore, always close the strip container
immediately after removing a test strip.
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......