Troubleshooting •
10
137
Errors and unusual behavior
without error messages
Some conditions may arise that have no associated error
or status message.
Message
Description
No message or unusual behavior
Meter display does not power on
■
Wait 10 seconds and try powering on the unit again.
■
Check that the meter has power.
• If you are using the external power adapter, is it
connected properly to the meter?
Or:
• Place the meter in the Handheld Base Unit.
Or:
• If you are using the Handheld Battery Pack, is it
properly installed in the meter?
Meter displays an unexpected result Refer to the package insert for the test strips.
Automatic shutdown
The meter powers off after a configurable time without
activity (e.g., pressing a key, touching the screen) to con-
serve energy. Reactivate the meter/screen as described
in the following:
Shutdown after time specified by
system administrator (default is 5 min-
utes, configurable by system adminis-
trator)
■
Press the On/Off button of the meter.
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......