9
• Maintenance and Care
132
How to perform maintenance
cleaning of the battery
compartment
1
With the meter powered off, press the battery com-
partment cover release tab and slide the cover off..
2
Use a lint-free cloth slightly dampened with water
to remove any residue from the areas highlighted in
blue in the illustration below (on the meter as well
as on the battery compartment cover).
3
Carefully dry the cleaned areas of the battery com-
partment cover and meter using a fresh lint-free
cloth.
Leave the parts disassembled to completely dry for at
least one hour.
You can either leave the batteries inserted in the battery
compartment cover or remove them when performing
the following maintenance cleaning steps.