Control Testing and Quality Control •
6
103
You hear a beep when you have applied enough control
solution (provided the beeper is enabled). The dropper
icon disappears and the test starts.
The result of the liquid quality control test is displayed. It
is automatically saved to memory.
The acceptable range of results for the liquid control is
displayed below the current result, along with
Pass
or
Fail
.
If a quality control test fails, an up arrow (too high) or
down arrow (too low) is displayed and flashes.
Note:
The arrow (next to the result) refers to the INR
result only.
If you have selected to display INR and %Quick or INR
and seconds, the (up or down) arrow next to the result
refers only to the INR value.
The printer icon only appears if the printer function
is activated. Otherwise it is not displayed.
QC Test
09:15 am
04/19/2012
0.9 INR
Fail:
110 %Q
04/19/2012 09:15 am
QC:
999
Code: 184
Op:
Operator 1
Par:
PT
Level: 1
(1.0 - 1.4 INR)
(53 - 99 %Q)
QC Test
09:15 am
04/19/2012
1.2 INR
78 %Q
04/19/2012 09:15 am
QC:
999
Code: 184
Op:
Operator 1
Par:
PT
Level: 1
(1.0 - 1.4 INR)
(53 - 99 %Q)
Pass:
QC Test
09:15 am
04/19/2012
QC:
999
Code: 184
Op:
Operator 1
Par:
PT
Level: 1
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......