2 About host interface features
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
44
Result requests
Result requests
In this section
About result requests (44)
Example of result query (46)
Example of unspecific result request (48)
Example of result request with ratio (collated result
request) (49)
About result requests
Due to the request/response principle, results are
transmitted to the host only upon the receipt of a Result
Request (BC09).
COBAS INTEGRA 400 plus analyzer and
Cedex Bio HT Analyzer allow requesting single results or
collated results. The BC09/LC10 thus contains a result
type selector.
q
We strongly advise
against
the use of collated
results, depending on the database size or fragmentation,
we cannot guaranty the instrument performance.
No specific result can be requested.
Only accepted results will be transmitted. (Auto accept
can be configured in
Configuration > Processing
> Result Handling Table
).
Usually each result will be transmitted only once. To re-
transmitted a result:
• On the instrument user interface, change to the result
workbench.
• Select the required results with the mouse.
• Select
Retransmit
in the
Process
menu.
We recommend using the following procedure and timing
for the result request loop: