8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
166
List of line codes for extended HIF
Line Code 61: Test Status and blocked status
Example
61_•3_•1<LF>
If the test status is blocked additional information is
provided regarding the reason of the impediment.
Description
Format Range
Range description
Line Code
61
Test Status
I2
16
Test disabled
31
Calibration postponed
3
Test blocked
17
OK
Blocked Status
I2
Detailed information is provided
below
y
Line Code 61: Test Status and blocked status
Status code
Meaning
0
Okay
1
No fluid in sample cup
2
Reagent not on board
3
Reagent is empty
4
No cassette definition found
5
Diluent not on board
6
Diluent is empty
7
No diluent definition found
8
No calibrator on board
9
Calibrator empty
10
No valid calibration data
11
No calibration defintion available
12
No lot values for calibration available
13
Last calibration is not accepted
14
Control not on board
15
Control empty
16
Control is not valid
17
No control definition available
18
No assigned values for control available
19
Last control not accepted
20
Repeat action not possible
21
Test disabled
22
Error in cuvette handling
23
Test aborted
24
Calculation not possible
25
No ISE Electrode
26
Activation not performed
27
ISE module not ready
28
Error in ISE module
y
Status codes