7
• Linearity Testing
92
7.2
Performing a linearity test
Overview of test procedure
A linearity test comprises the following steps:
■
Check the lot number of the linearity solutions.
■
Check the lot number of the test strips.
■
Perform the test with a minimum of three linearity
solutions.
Starting a linearity test
The following description assumes that the meter is
powered on and the
Main Menu
is displayed.
1
Touch
in the
Main Menu
to open the
Main
Menu 2
screen.
2
Touch
Linearity
to start the linearity test. The
Line-
arity Test
screen opens.
Linearity Test
12345678?
Use Linearity Lot
12:48 pm
09/18/13
Main Menu
Review Results
Control Test
Patient Test
12:48 pm
09/18/13
Main Menu 2
Maintenance
Strip Lots
Profi ciency
Control Lots
Beeper
Linearity
Diagnostics
Linearity Lots
Admin.