Example
Understanding Control Test Results
Example
Example
ce Center 1-800-858-8072
hlighted area.
ntrol Solution
the meter.
may not be
6, using Level 2
al pharmacy.
The label on your test strip vial shows the acceptable ranges for
Level 1 and Level 2 test solutions. Your control test results should
be within the listed range. Make sure you compare the result to the
range for the level of control solution you used (Level 1 or Level 2).
When a control test result is within the acceptable range,
your test strips and meter are working properly.
“
Understanding
Control Test
Results
prints
100m, 43y, 18k.
Gradient blend
behind range
chart prints 5m,
20y, 1k to white.
Range (mg/dL)
Level 1 (LO)
42–72
Level 2 (HI)
290–328