씰
If you performed the test with Accu-Chek Active Control 1,
refer to row 1 for the permitted range.
씰
If you performed the test with Accu-Chek Active Control 2,
refer to row 2 for the permitted range.
If the result is within the stated range, you need to still carry out a
visual plausibility test. It is important that this check be performed
within 30–60 seconds after control solution was applied. Any later
than this and comparison is no longer possible owing to excessive
discolouration of the test strip.
If the result is outside the stated range, perform a second quality
control test. If the second result is still outside the range, please call
your customer support and service centre.
씰
Pull the strip out of the meter. The result is saved as a control
reading (which is ignored when the averages are calculated),
and the meter switches off.
93
man_04493168001_02_EnUK.indd 93
1/12/2006, 11:29:15 ðì
Summary of Contents for ACCU-CHECK Active
Page 1: ...ACCU CHEK Active BLOOD GLUCOSE SYSTEM Reference Manual...
Page 6: ...Parts summary Accu Chek Active 8 7 1 3 2 9...
Page 8: ...Parts summary Accu Chek Active 10...
Page 9: ...Coding chip slot Infrared interface 10 11 11...
Page 24: ...3 2 Basic settings...
Page 26: ...3 2 Basic settings...
Page 28: ...3 2 Basic settings...
Page 30: ...3 2 Basic settings...
Page 32: ...3 2 Basic settings...
Page 36: ...3 3 Coding...
Page 40: ...4 3 Performing a blood glucose test...
Page 42: ...4 3 Performing a blood glucose test...
Page 44: ...4 3 Performing a blood glucose test...
Page 48: ...4 3 Performing a blood glucose test...
Page 50: ...4 3 Performing a blood glucose test...
Page 52: ...4 3 Performing a blood glucose test...
Page 56: ...4 3 Performing a blood glucose test...
Page 60: ...4 3 Performing a blood glucose test...
Page 64: ...5 2 Retrieving results...
Page 66: ...5 2 Retrieving results...
Page 68: ...5 2 Retrieving results...
Page 70: ...5 2 Retrieving results...
Page 72: ...5 2 Retrieving results longer than 3 seconds...
Page 74: ...5 2 Retrieving results longer than 3 seconds longer than 5 seconds...
Page 78: ...5 3 Downloading data to a PC longer than 3 seconds...
Page 80: ...5 3 Downloading data to a PC...
Page 84: ...6 2 Preparing for the quality control test...
Page 88: ...6 3 Quality control test procedure...
Page 90: ...6 3 Quality control test procedure...
Page 92: ...6 3 Quality control test procedure...
Page 96: ...6 3 Quality control test procedure...
Page 98: ...6 4 Performing a full display check longer than 3 seconds...
Page 102: ...7 1 Cleaning your Accu Chek Active meter...
Page 104: ...7 1 Cleaning your Accu Chek Active meter...
Page 108: ...7 2 Battery life and battery replacement...
Page 127: ...A B D C E F H I G...