background image

INSTALLATION DATA

2374-510

UNIVERSAL AIR PRESSURE 

SENSING SWITCH

191 E. North Avenue
Carol Stream  Illinois 60188 USA
Customer Service Telephone 1.800.304.6563
Customer Service Facsimile 1.800.426.0804
[email protected]

Invensys™ and Robertshaw

®

 are trademarks of Invensys 

plc., its subsidiaries and/or affiliated companies. All 
other brands mentioned may be the trademarks of their 
respective owners.

www.Uni-Line.com
www.InvensysControls.com
©2011 Invensys Controls
5/11 - LTNS2000008 00

For Technical Service
Telephone 1.800.445.8299
Facsimile 1.630.260.7294
[email protected]

Carton Contents

(1) field adjustable switch
(1) black set point adjustment screw
(1) hexagon key wrench 
(5) color-coded springs
(8) color-coded orifices
(1) half strap mounting bracket
(1) angle ("L") bracket
(1) mounting screw for optional brackets
(2) #6 x ⅜" mounting screws for optional brackets

General DesCription & operation

The Robertshaw

®

 2374-510 Air Pressure Sensing Switch 

has  a  glass-filled  polycarbonate  housing  containing  a 

sensing  diaphragm  and  an  integral  snap-acting  switch 

with three male 90° quick-connect terminals. The switch 

can sense positive, negative or differential air pressure. 

The  field  adjustable  set  point  range  of  this  switch  is 

0.10" W.C. to 10.0" W.C. Using the switch accessories 

contained  in  this  kit,  the  switch  can  be  applied  to  a 

wide  variety  of  residential  and  light  commercial  HVAC 

applications.

set point ranGe & aDjustment

Calibration requires a manometer (not included in the kit) 

as well as the included 

7

32

" hex wrench.

1. Establish the set point as follows. Referring to Table 2 in 

this manual, select the appropriate spring for the required 

set point range. Insert the spring into the center well of 

the mounting pan (light gray side of the switch housing).
2. Insert the black set point adjustment screw, and rotate 

it manually until the threads are engaged.
3. Connect the switch to a manometer. Using  the 

7

32

hex wrench provided in the kit, turn the adjustment screw 

in small increments until the desired set point is reached. 

Turn  the  screw  clockwise  to  increase  the  set  point  or 

counterclockwise to decrease the set point. For precise 

calibration,  confirm  the  set  point  at  actual  operating 

temperature  with  a  manometer.  Following  precise 

calibration, if desired, seal the adjusting screw using a 

UV curable adhesive/sealant. 

Do not place the switch 

in operation without knowing what the set point is: 

doing so could create a hazardous situation.

 

mountinG

Using the (2) #6 x ⅜" mounting screws included in the 

kit, mount via the integral foot bracket (see Fig. 1) 

or

, via 

either of the optional brackets (see Figs 4 & 5) included 

in the kit. If using one of the optional brackets, attach it 

to the switch with the self-tapping screw provided before 

mounting  the  switch.  Select  a  mounting  location  free 

from vibration. Mount with the diaphragm in any vertical 

plane. Avoid mounting with the sample line connections 

directed upward.

eleCtriCal ConneCtions

The snap switch has three ¼" 90° male quick connect 

terminals. Before pressure is applied to the diaphragm, 

the  switch  contacts  are  in  the  deactivated  position  as 

shown in Figure 2.

air samplinG ConneCtion

Integral sample line connectors, located on both sides 

of  the  diaphragm,  accept 

3

16

"  ID  flexible  tubing.  See 

Figure 3. The High or Positive inlet (P1) is black and the 

Low or Negative inlet (P2) is gray. Connect the sample 

lines as follows:

positive pressure only:

 Connect the sample line to P1; 

P2 remains open to the atmosphere. 

negative pressure only:

 Connect the sample line to P2; 

P1 remains open to the atmosphere.

two negative samples:

 Connect higher negative sample 

to P2; lower sample to P1.

two positive samples:

 Connect higher positive sample 

to P1; lower sample to P2.

one  positive  and  one  negative:

  Connect  positive 

sample to P1; connect negative sample to P2.

usinG a Flow-restriCtinG oriFiCe

Some  applications  require  a  delayed  switching  action 

after  set  point  is  reached.  The  delay  is  created  by 

inserting an orifice plug into either or both of  the sample 

line connectors to restrict air flow. Eight orifice plugs in 

four color-coded sizes are included in the kit as shown 

in Table 3. Note that the measuring device and the air 

switch must both contain the same size restricting orifice in 

order to obtain an accurate measurement of the set point. 

Reviews: