background image

Table section 4 / test gas analyses 
 
 

 
 
PGC9300 Maintenance Logbook  ·  EN07  ·  2018, June 26th 

55 

 

 

OK? 

Verification of the chromatogram of the internal calibration gas 

 

Verification of the chromatogram of test gas 1 

 

Verification of the chromatogram of test gas 2 

 

Verification of the chromatogram of test gas 3 

 

Verification of the chromatogram of test gas 4 

 

Verification of the chromatogram of test gas 5 

 

Verification of the chromatogram of test gas 6 

 

Checking the range of the current outputs 

 

Verification of the fiscal metering values transmitted at the outputs 

 

Verification of the software and kernel CRCs 

 

Are all results correctly logged? 

 

 
 

Remarks 

 

 

 

 

 

 

 

 

 

 

 

Signatures 

 

The signatures relate to the entire 
metrological check / recalibration, 
i.e. including the entries  
on pages 51 to 55. 

Service 

 

 

Authority 

 

 

 
 

Summary of Contents for PGC 9300

Page 1: ...Process Gas Chromatograph PGC 9300 MAINTENANCE LOGBOOK Reliable Measurement of Gas Status 2018 JUNE 26TH Version 07...

Page 2: ...ment may serve as reference for translations into other languages Please use in case of any uncertainties the German version as main reference Note Unfortunately paper is not updated automatically whe...

Page 3: ...PGC year of manufacture Meter serial number GC 9300 manufacturer number GC 9300 year of manufacture GC 9300 software version Approved for Calorific value Gas quality Operator Station Commissioning Typ...

Page 4: ...1 1 1 2 Cylinder pressures 2 1 2 Maintenance by service technicians 2 2 Manufacturer s requirements 3 2 1 Annual maintenance of the process gas chromatograph 3 2 2 Metrological check re calibration 3...

Page 5: ...ion 1 of this maintenance logbook must be documented during every inspection Routine maintenance should be performed once a year by qualified personnel according to the manufacturer s instructions and...

Page 6: ...ould be checked for Carrier gas 1 Carrier gas 2 if applicable Internal calibration gas 1 2 Maintenance by service technicians This section records all changes and measures carried out by service techn...

Page 7: ...s with the opening of the calibration lock and the implementation of a basic calibration select tab Detail 01 GC 9300 Mode Basic Calibration Following basic calibration the new response factors tab De...

Page 8: ...Table section 1 PGC9300 Maintenance Logbook EN07 2018 June 26th 4 Table Section 1 Cylinder pressures To be performed at Every inspection Every maintenance Entries made by Operator RMG Service...

Page 9: ...ction 1 cylinder pressures PGC9300 Maintenance Logbook EN07 2018 June 26th 5 Cylinder changed Device faulty Carriergas I Int calibration gas Carrier gas II Date Initials Pressure bar Pressure bar Pres...

Page 10: ...ction 1 cylinder pressures PGC9300 Maintenance Logbook EN07 2018 June 26th 6 Cylinder changed Device faulty Carriergas I Int calibration gas Carrier gas II Date Initials Pressure bar Pressure bar Pres...

Page 11: ...ction 1 cylinder pressures PGC9300 Maintenance Logbook EN07 2018 June 26th 7 Cylinder changed Device faulty Carriergas I Int calibration gas Carrier gas II Date Initials Pressure bar Pressure bar Pres...

Page 12: ...nance Logbook EN07 2018 June 26th 8 Table section 2 Maintenance measures general To be performed at Annual routine maintenance Software update Metrological check Re calibration Repair Entries made by...

Page 13: ...k EN07 2018 June 26th 9 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Page 14: ...EN07 2018 June 26th 10 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Page 15: ...EN07 2018 June 26th 11 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Page 16: ...EN07 2018 June 26th 12 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Page 17: ...EN07 2018 June 26th 13 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Page 18: ...EN07 2018 June 26th 14 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Page 19: ...EN07 2018 June 26th 15 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Page 20: ...EN07 2018 June 26th 16 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Page 21: ...EN07 2018 June 26th 17 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Page 22: ...EN07 2018 June 26th 18 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Page 23: ...able section 3 PGC9300 Maintenance Logbook EN07 2018 June 26th 19 Table section 3 Annual routine maintenance To be performed at Annual routine maintenance Entries made by Trained personnel RMG Service...

Page 24: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Page 25: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Page 26: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Page 27: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Page 28: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Page 29: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Page 30: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Page 31: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Page 32: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Page 33: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Page 34: ...Table section 4 PGC9300 Maintenance Logbook EN07 2018 June 26th 30 Table section 4 Test gas analyses To be performed at Metrological check Re calibration Entries made by Trained personnel RMG Service...

Page 35: ...nalysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Verific...

Page 36: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 37: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 38: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 39: ...gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verificati...

Page 40: ...2 Analysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Veri...

Page 41: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 42: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 43: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 44: ...est gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verific...

Page 45: ...nalysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Verific...

Page 46: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 47: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 48: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 49: ...gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verificati...

Page 50: ...2 Analysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Veri...

Page 51: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 52: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 53: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 54: ...est gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verific...

Page 55: ...nalysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Verific...

Page 56: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 57: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 58: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 59: ...gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verificati...

Page 60: ...2 Analysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Veri...

Page 61: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 62: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 63: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 64: ...est gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verific...

Page 65: ...nalysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Verific...

Page 66: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 67: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 68: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 69: ...gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verificati...

Page 70: ...2 Analysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Veri...

Page 71: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 72: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 73: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 74: ...est gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verific...

Page 75: ...nalysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Verific...

Page 76: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 77: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 78: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 79: ...gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verificati...

Page 80: ...2 Analysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Veri...

Page 81: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 82: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Page 83: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Page 84: ...est gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verific...

Page 85: ...For further information please visit our website www rmg com or contact your local sales support office to learn more about the RMG products RMG Messtechnik GmbH Otto Hahn Stra e 5 35510 Butzbach Ger...

Reviews: