Device description
The analyzer consists of the following components: The controller with graphic display and control
panel is on the left side of the device. 4 cable bushings are provided on the lower controller
housing. The dosing pump is on the right side and the measuring chamber is below the dosing
pump.
The dosing pump is plugged onto the motor shaft of the motor and snaps into the receptacle. It
can be removed without tools. The measuring chamber is attached to 2 guide pins on the
controller housing. This can also be removed without tools and is undone with 2 captive locking
pins. Dosing plugs, actuator plugs, inlet- and outlet plugs are attached to the measuring chamber
in the same way and can be quickly undone.
The measuring chamber, which is always depressurised and always filled to prevent germs, is in the
middle. The actuator plug with a white high-power LED is in the middle of the measuring chamber.
The sensor system is located in the housing. The agitator (magnetic agitator), which is firmly
connected to the housing, is below the measuring chamber.
The water connections for hoses with an outer diameter of 6 mm are below the reagent bottle.
Inlet with sample water (left) and outlet (right). Both are firmly bolted to the wall bracket.
There is a solenoid valve in the supply line behind the reagent bottle.
Figure: View of the solenoid valve (no reagent bottle installed)
For a more detailed description of the components, see page 30.
SYCON 2702 in detail