65-2437RKSS-05 CO Detector • 7

9.

If using shielded cable, connect the cable’s drain wire to an available chassis ground at 
the controller. RKI controllers typically have a ground stud that is a convenient 
grounding location.

10. Reinstall the junction box cover.

Start Up

This section describes procedures to start up the CO detector and place the detector into 
normal operation.

Introducing Incoming Power

1.

Complete the installation procedures described earlier in this manual.

2.

Verify that the power wiring to the controller is correct and secure. Refer to the 
controller operator’s manual.

3.

Turn on power to the controller.

4.

Turn on the controller.

5.

Verify that the controller is on and operating properly. Refer to the controller 
operator’s manual.

CAUTION:

Allow the detector to warm up for 5 minutes before you continue with the next 
section, “Setting the Zero (Fresh Air) Reading”.

Setting the Zero (Fresh Air) Reading

WARNING: If the detector installation is explosion proof, do not remove the detector 

housing cap or junction box cover while the circuits are energized unless the 
area is determined to be non-hazardous. Keep the detector housing cap and 
junction box cover tightly closed during operation.

CAUTION:

If you suspect the presence of CO in the monitoring environment, use the calibration 
kit and a zero air calibration cylinder to introduce “fresh air” to the CO detector and 
verify an accurate zero setting. See “Calibration” on page 12 for instructions to 
introduce zero air to the CO detector.

1.

Verify that the detector is in a fresh air environment (environment known to be free of 
carbon monoxide, other toxic and combustible gases, and of normal oxygen content, 
20.9%).

2.

Verify a reading of 0 ppm CO at the controller.

If the display reading is 0 ppm CO, start up is complete. The detector is in normal 
operation. If the display reading is not 0 ppm CO, continue with step 3.

3.

Perform a zeroing operation at the controller. See the controller operator’s manual for 
directions.

Summary of Contents for 65-2437RKSS-05

Page 1: ...RKI Instruments Inc www rkiinstruments com 65 2437RKSS 05 Carbon Monoxide Detector Operator s Manual Part Number 71 0439 Revision P1 Released 9 14 17...

Page 2: ...calibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type...

Page 3: ...t costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRES...

Page 4: ...he CO Detector 4 Wiring the CO Detector to a Controller 5 Startup 7 Introducing Incoming Power 7 Setting the Zero Signal 7 Maintenance 8 Preventive Maintenance 8 Troubleshooting 9 Replacing Components...

Page 5: ...e instructions and warnings in this manual to assure proper and safe operation of the 65 2437RKSS 05 and to minimize the risk of personal injury Be sure to maintain and periodically calibrate the 65 2...

Page 6: ...on describes the components of the 65 2437RKSS 05 CO detector The 65 2437RKSS 05 consists of the 65 2438 CO detector and junction box Figure 1 CO Detector Component Location Stainless Steel J Box Lock...

Page 7: ...ntly bonded to the cap Two wires extend from the top of the detector housing body Use these wires to connect the CO detector to the junction box terminal strip The housing includes a four socket patte...

Page 8: ...the junction box A locking set screw on the junction box cover allows you to secure the junction box cover and prevent it from being removed Installation This section describes procedures to mount th...

Page 9: ...source is OFF before you make wiring connections 1 Turn off the controller 2 Turn off or unplug power to the controller 3 Remove the junction box cover WARNING To maintain the explosion proof classif...

Page 10: ...junction box cover to the junction box 7 Route the cable or wires leading from the CO detector through one of the conduit hubs at the controller housing Use appropriate conduit fittings and constructi...

Page 11: ...the Zero Fresh Air Reading WARNING If the detector installation is explosion proof do not remove the detector housing cap or junction box cover while the circuits are energized unless the area is dete...

Page 12: ...ble external alarms 2 Verify that the controller display reading for the channel you are testing is 0 If the display reading is not zero set the zero reading of the detector as described in the Start...

Page 13: ...lt or Unable to Calibrate Detector responds slowly or does not respond during response test Unable to accurately set the zero or response reading during calibration Detector requires frequent calibrat...

Page 14: ...in the detector housing 8 Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body 9 Turn on or plug in power to the controller 10 Turn on the contro...

Page 15: ...ect 8 Reinstall the junction box cover 9 Turn on or plug in power to the controller 10 Turn on the controller and place into normal operation CAUTION Allow the replacement detector to warm up for 5 mi...

Page 16: ...he fixed flow regulator to the calibration cup 4 Put the controller into its calibration program Setting the Zero Reading NOTE If you can verify that the CO detector is in a fresh air environment you...

Page 17: ...regulator from the cylinder 8 Remove the calibration cup from the detector NOTE For convenience leave regulator and calibration cup connected by the sample tubing 9 Allow about 45 seconds for the gas...

Page 18: ...9 65 2437RKSS 05 CO Detector Operator s Manual this document 81 0064RK Calibration cylinder 50 PPM CO in air 17 liter 81 0064RK 01 Calibration cylinder 50 PPM CO in air 34 liter steel 81 0064RK 03 Cal...

Reviews: