61-1000RK-05/61-0140RK-05 Combustible Gas Detector • 11
Maintenance
This section describes maintenance procedures. It includes preventive maintenance,
troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum
performance of the combustible gas detector. It includes daily and quarterly procedures.
Daily
Verify a display reading of 0% LEL at the controller. Investigate significant changes in the
reading.
Quarterly
Calibrate the detector as described in the Calibration section of this manual.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the combustible gas detector.
NOTE:
This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.
Table 2:Troubleshooting the Combustible Gas Detector
Condition
Symptom(s)
Probable Causes
Recommended Action
Fail Condition
• Controller indicates a
fail condition.
• The detector wiring is
disconnected or
misconnected.
• The detector zero
signal is low enough to
cause a fail condition.
• The detector is
malfunctioning.
1. Verify that the detector wiring is
correct and secure.
2. Calibrate the detector.
3. If the fail condition continues, replace
the detector.
4. If the fail condition continues, contact
RKI for further instruction.
Slow or No
Response/
Difficult or
Unable to
Calibrate
• Unable to accurately
set the zero or
response reading
during calibration.
• Detector requires
frequent calibration.
Note:
Under “normal”
circumstances, the
detector requires
calibration once every
three months.
Some applications may
require a more frequent
calibration schedule.
• The calibration cylinder
is low, out-dated, or
defective.
• The calibration gas is
not an appropriate
concentration.
• The detector is
malfunctioning.
1. Verify that the calibration cylinder
contains an adequate supply of a
fresh test sample.
2. Verify that the calibration gas
concentration is appropriate for the
detector.
3. If the calibration/response difficulties
continue, replace the detector.
4. If the calibration/response difficulties
continue, contact RKI for further
instruction.