AirLink T3A Operator’s Manual
Maintenance
•
63
Replacing the Fuse
WARNING: Do not remove the sensor housing cap or enclosure lid while the circuits are
energized unless the area is determined to be non-hazardous. Keep the sensor
housing cap and enclosure lid tightly closed during operation.
CAUTION:
The internal components can be static sensitive. Use caution when opening the
enclosure and handling internal components. DO NOT use any metal objects or
tools to remove the fuses.
1. Disconnect or turn off power to the AirLink T3A.
2. Unscrew the enclosure lid and set it aside.
3. Grab the thumbscrews and gently lift the internal system out of the enclosure. It can rest on
the edge of the enclosure.
NOTE:
Disconnecting the sensor connector plug from the sensor housing will allow
for the complete removal of the internal system from the device enclosure.
Disconnecting the internal system may provide ease in accessing the control
board terminals for wiring. Reconnect the sensor connector plug before
reinstalling the internal system.
4. Locate the fuse that needs to be replaced.
5. Pull the old fuse straight out.
6. Align the pins on the replacement fuse with the sockets in the fuse holder.