Operation Settings and Calibration
AirLink T3A Operator’s Manual
3. Press the ADD button to select “Yes” to reset the Zero and Calibration values and to advance
to the Reset Zero & Cal Only confirmation screen. If you do not wish to reset the zero and
calibration values, press the SUB button to select “No” to leave the Product Settings and
Configuration menu and to return the device to normal operating mode.
4. If “Yes” is selected to reset the zero and calibration values:
5. Press the ADD button to select “Yes” to confirm that you want to reset the Zero and
Calibration values and to return the device to normal operating mode. If you do not wish to
continue to reset the Zero and Calibration values, press the SUB button to select “No” to leave
the Product Settings and Configuration menu and to return the device to normal operating
mode.
6. Press and release the MENU button to return to Normal Operating Mode.
NOTE:
If the AirLink T3A’s stored Zero and Calibration values are reset, the device
MUST be zeroed and calibrated for proper and safe operation of the device.
Operation Settings and Calibration
The Operation Settings menu allows you to:
• perform a zero adjust
• perform a span adjust
• change sensor alarm settings
• change sensor radio address
NOTE:
After 5 minutes of no interaction with the device, the unit will automatically return to
normal operating mode.
NOTE:
The alarms will NOT activate, even in the presence of gas, until you have been out of
the Operation Settings menu for 1 minute.
The AirLink T3A continues monitoring for gas while in the Operation Settings menu.