Turning on a SDM-6000 System • 90
8 . If the GX-6000 is equipped with a Li-ion battery pack that needs to be charged,
the Status LED will begin to flash orange. If the battery pack is fully charged,
the Status LED will be solid orange. See “Charging a GX-6000 in a Docking
Station” on page 105 for a complete description of charging the GX-6000.
9 . Push the red gas in/gas out cover down until it touches the GX-6000 and then
release it. It should lock into place against the top of the GX-6000. If it returns
to its original position, attempt to lock it in place again.
10 . The GX-6000 will turn on once the gas in/gas out cover is in place. Any
instrument charging that was taking place will stop and the SDM-6000 Status
LED will blink green.
11 . Repeat step 7 - step 10 for any additional SDM-6000s and GX-6000s.
12 . The GX-6000 will begin its power up sequence. The status of the instrument is
shown by the Station/Instrument Indicators at the bottom of the Function Tab.
The color of the corresponding indicator and the logo shown in it will indicate
the status of the instrument. For example, the master station screen below
indicates that the instrument in position 1 is connected to the SDM-6000 and
the Active box in the Unit Tab is selected. It also indicates that the instrument
in position 2 is connected to the SDM-6000, the Active box in the Unit Tab is
selected, and the GX-6000 is due for calibration. See “Function Tab” on page 8
for a complete description of the indicators’ colors and logos.