RP-2009 Pump for the GX-2009

  13

Alarms

The table below outlines the alarms that may occur on the RP-2009.

Maintenance

Replacing The Battery

1. Locate the black plastic battery cap on one end of the pump unit.

2. Push down and rotate battery cap counterclockwise until it stops, 1/4 

of a turn.

3. Pull the battery cap away from the pump unit.

4. Remove the old battery from the pump unit. It will slide right out.

5. Insert the new b end first into the pump unit. Use an AA 

alkaline. The battery orientation is shown on the bottom of the pump 
unit.

6. Slide the battery cap over the battery onto the pump unit, push it down 

and turn it clockwise unit is stops beneath the instrument release tab.

Replacing the Cotton Dust Filter

1. Locate the clear plastic filter holder at one end of the pump unit. 

2. Grasp the ridged sides of the filter holder and turn it 1/4 of a turn 

counterclockwise.

Alarm Type

Visual 

Indications

Audible 

Indications

Other 

Indications

Recommended 

Action

Low Battery

POWER 

LED flashing

none

none

Replace the battery.

Dead Battery

POWER 

LED flashing

constant 

buzzer

none

Replace the battery.

Low Flow

NOTE: A low flow alarm 
may get combined with a 
low or dead battery 
alarm since a low battery 
voltage may cause the 
pump to stop working.

LOW FLOW 

LED flashing

intermittent 

buzzer

pump stops

• Check the rubber 

sample nipple and 

filters to make sure 

they are clear.

• If a hose and probe 

are used, check 

them for kinks or 

obstructions.

• Press and release 

the POWER button 

to reset the alarm.

• If the low flow 

condition persists, 

the pump may need 

servicing.

Summary of Contents for RP-2009

Page 1: ...www rkiinstruments com RP 2009 Pump for the GX 2009 Operator s Manual Part Number 71 0267RK Revision P4 Released 10 26 18...

Page 2: ...ent costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPR...

Page 3: ...3 Replacing the Hydrophobic Disk Filter and Wire Mesh Disk 14 Parts List 15 WARNING Understand manual before operating and for proper battery type Substitution of components may impair intrinsic safet...

Page 4: ...able 1 lists specifications for the RP 2009 pump Table 1 Specifications Power Supply 1 ea AA size alkaline Operating Time 8 hours minimum at 25 C Operating Temperature Humidity 20 C to 50 C 0 85 Relat...

Page 5: ...e 1 RP 2009 Pump Component Location Top Power Button Low Flow LED Power LED Instrument Release Indicator Sensor Adapter Inlet Fitting GX 7 Type Back Front Buzzer Opening Instrument Lock Indicator Pump...

Page 6: ...pump unit on and off The buzzer opening allows the buzzer s sound to emanate from the buzzer The power LED is on when the pump unit is on The low flow LED flashes when the pump is experiencing a low f...

Page 7: ...wn on the RP 2009 s battery cap and rotate it 1 8 turn counterclockwise so that the indicator tab on the battery cap is next to RELEASE Figure 2 Releasing the Battery Cap WARNING Do not install the GX...

Page 8: ...GX 2009 Be sure that the RP 2009 is fully engaged on the instrument When fully engaged the GX 2009 should not move around at all CAUTION If the RP 2009 is not fully engaged the unit will not respond...

Page 9: ...K Figure 4 Locking the Battery Cap 6 Attach the rubber sample nipple or an optional hose and probe to the inlet fitting Figure 5 Connecting the Optional Hose Probe 7 Attach the RP 2009 GX 2009 assembl...

Page 10: ...GX 2009 If an optional hose and probe are used instead of the rubber sample nipple hold the probe in the area to be sampled To sample from a well or other similar sample area the hose and probe may be...

Page 11: ...at the indicator tab on the battery cap is next to RELEASE Figure 6 Releasing the Battery Cap WARNING Do not remove the GX 2009 from the RP 2009 without rotating the battery cap to RELEASE Removing th...

Page 12: ...16 Press the bottom of the release tab toward the pump unit and pull up on the GX 2009 Figure 7 GX 2009 Removal LOW FLOW POWER RESET Indicator Tab LOCK RELEASE OPEN AIR POWER MODE H2S CO LEL O2 RP 200...

Page 13: ...lease tab Replacing the Cotton Dust Filter 1 Locate the clear plastic filter holder at one end of the pump unit 2 Grasp the ridged sides of the filter holder and turn it 1 4 of a turn counterclockwise...

Page 14: ...d of the pump unit 2 Grasp the ridged sides of the filter holder and turn it 1 4 of a turn counterclockwise 3 Pull the filter holder away from the pump unit Do not lose the cotton dust filter 4 The hy...

Page 15: ...it snaps into place Parts List Table 2 lists replacement parts and accessories for the RP 2009 pump Table 2 Parts List Part Number Description 07 2005RK Gasket filter screen retaining 07 6027RK O rin...

Reviews: