background image

 

www.rkiinstruments.com

RP-2009 Pump for the GX-2009

Operator’s Manual

Part Number: 71-0267RK

Revision: P4

Released: 10/26/18

Summary of Contents for RP-2009

Page 1: ...www rkiinstruments com RP 2009 Pump for the GX 2009 Operator s Manual Part Number 71 0267RK Revision P4 Released 10 26 18...

Page 2: ...ent costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPR...

Page 3: ...3 Replacing the Hydrophobic Disk Filter and Wire Mesh Disk 14 Parts List 15 WARNING Understand manual before operating and for proper battery type Substitution of components may impair intrinsic safet...

Page 4: ...able 1 lists specifications for the RP 2009 pump Table 1 Specifications Power Supply 1 ea AA size alkaline Operating Time 8 hours minimum at 25 C Operating Temperature Humidity 20 C to 50 C 0 85 Relat...

Page 5: ...e 1 RP 2009 Pump Component Location Top Power Button Low Flow LED Power LED Instrument Release Indicator Sensor Adapter Inlet Fitting GX 7 Type Back Front Buzzer Opening Instrument Lock Indicator Pump...

Page 6: ...pump unit on and off The buzzer opening allows the buzzer s sound to emanate from the buzzer The power LED is on when the pump unit is on The low flow LED flashes when the pump is experiencing a low f...

Page 7: ...wn on the RP 2009 s battery cap and rotate it 1 8 turn counterclockwise so that the indicator tab on the battery cap is next to RELEASE Figure 2 Releasing the Battery Cap WARNING Do not install the GX...

Page 8: ...GX 2009 Be sure that the RP 2009 is fully engaged on the instrument When fully engaged the GX 2009 should not move around at all CAUTION If the RP 2009 is not fully engaged the unit will not respond...

Page 9: ...K Figure 4 Locking the Battery Cap 6 Attach the rubber sample nipple or an optional hose and probe to the inlet fitting Figure 5 Connecting the Optional Hose Probe 7 Attach the RP 2009 GX 2009 assembl...

Page 10: ...GX 2009 If an optional hose and probe are used instead of the rubber sample nipple hold the probe in the area to be sampled To sample from a well or other similar sample area the hose and probe may be...

Page 11: ...at the indicator tab on the battery cap is next to RELEASE Figure 6 Releasing the Battery Cap WARNING Do not remove the GX 2009 from the RP 2009 without rotating the battery cap to RELEASE Removing th...

Page 12: ...16 Press the bottom of the release tab toward the pump unit and pull up on the GX 2009 Figure 7 GX 2009 Removal LOW FLOW POWER RESET Indicator Tab LOCK RELEASE OPEN AIR POWER MODE H2S CO LEL O2 RP 200...

Page 13: ...lease tab Replacing the Cotton Dust Filter 1 Locate the clear plastic filter holder at one end of the pump unit 2 Grasp the ridged sides of the filter holder and turn it 1 4 of a turn counterclockwise...

Page 14: ...d of the pump unit 2 Grasp the ridged sides of the filter holder and turn it 1 4 of a turn counterclockwise 3 Pull the filter holder away from the pump unit Do not lose the cotton dust filter 4 The hy...

Page 15: ...it snaps into place Parts List Table 2 lists replacement parts and accessories for the RP 2009 pump Table 2 Parts List Part Number Description 07 2005RK Gasket filter screen retaining 07 6027RK O rin...

Reviews: