Chapter 5: User Mode and Calibration
OX-04 Operator’s Manual
Chapter 5: User Mode and Calibration
Overview
This section describes the OX-04 in User Mode. See Table 8 below for a list of the items found in
User Mode, the page that the item’s instructions can be found on, and a short description of the
item.
Table 8: User Mode Menu Items
User Mode
Menu Item
Description
BUMP (page 41)
Perform a bump test.
BUMP
Perform a bump test.
START
Begin the warmup sequence and enter Measuring Mode.
ESCAPE
Return to the
BUMP
menu item.
GAS CAL (page 43)
Perform a fresh air adjustment, perform a span adjustment, change the calibration
gas concentration.
AIR (page
45)
Perform a fresh air adjustment.
A-CAL
(page 46) or
E-CAL (page
50) depend
-
ing on
E-
CAL
User
Mode set
-
ting*
A-CAL (or
E-CAL)
Perform a span adjustment.
START
Begin the warmup sequence and enter Measuring
Mode.
CAL-P
Set the calibration gas concentration.
ESCAPE
Return to the
A-CAL
item in the
GAS CAL
menu.
ESCAPE
Return to the
GAS CAL
item in User Mode.
CAL SET (page 54)
Change parameters related to calibration.
CAL.RMDR
(page 54)
ON
(factory setting): The instrument notifies the user upon startup
when a calibration is due. Notification type depends on
CAL.EXPD
setting below.
CAL.DATA
screen appears in Display
Mode.
OFF
: No notification upon startup when a calibration is due.
CAL.DATA
screen does not appear in Display Mode.