54
•
Chapter 5: User Mode and Calibration
OX-04 Operator’s Manual
10. Use AIR to scroll to
ESCAPE
.
11. Press and release POWER MODE to return to the
CAL--P
item in the
A-CAL
menu.
12. Press AIR to scroll to
ESCAPE
.
13. Press and release POWER MODE to return to the
A-CAL
item in the
GAS CAL
menu.
14. See “Exiting the GAS CAL Menu” on page 54 to return to User Mode.
Exiting the GAS CAL Menu
1. While in the
GAS CAL
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER MODE to return to the
GAS CAL
item in User Mode.
3. See “Entering Measuring Mode (START)” on page 69 to enter Measuring Mode.
Setting Calibration Parameters (CAL SET)
The
CAL SET
menu has 4 items:
CAL.RMDR
,
CAL.INT
,
CAL.EXPD
, and
ESCAPE
.
1. While in User Mode, press AIR to scroll to
CAL SET
.
2. Press and release POWER MODE. The
CAL.RMDR
item appears.
CAL.RMDR
ON
(factory setting): The OX-04 gives an indicator at start up if it is due for calibration. The type
of indicator depends on the
CAL.EXPD
setting (see page 55).
OFF
: The OX-04 does not give an indicator at start up if it is due for calibration.
1. After entering the
CAL SET
menu, press AIR to scroll to
CAL.RMDR
.
2. Press and release POWER MODE. The current setting flashes.
CAL SET
CAL .RMDR
CAL .RMDR