GX-3R Pro Operator’s Manual
Chapter 5: User Mode and Calibration
•
98
Adjusting Man Down and Panic Settings (MAN
DOWN)
The
MAN DOWN
menu item has 4 items:
MAN DOWN
,
PANIC
,
MAN DOWN TIME
, and
ESCAPE
.
1. While in User Mode, use AIR to place the cursor next to
MAN DOWN
.
2. Press and release POWER MODE. The
MAN DOWN
menu appears.
Turning Man Down On/Off
ON
: The Man Down alarm is triggered if the instrument detects no motion for the period of time
defined in
WARNING 1 TIME
,
WARNING 2 TIME
, and
ALARM TIME
below.
OFF
(factory setting): The Man Down alarm is not triggered.
1. While in the
MAN DOWN
menu, use AIR to place the cursor next to
MAN DOWN
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
MAN DOWN
menu.
5. See “ESCAPE” on page 101 to return to User Mode.
MAINTE
7:49
BUMP SETTING
MAN DOWN
ALARM SETTING
USER MODE
MAINTE
7:49
MAN DOWN
PANIC
MAN DOWN TIME
MAN DOWN MENU
MAINTE
7:49
MAN DOWN
PANIC
MAN DOWN TIME
MAN DOWN MENU