GX-3R Pro Operator’s Manual
Appendix C: Gas Select Mode
•
192
7. Press and release POWER MODE to save the setting.
8. Repeat Step 1 - Step 7 to change the target gas for other channels.
9. Use AIR to scroll to
ESCAPE
.
10. Press and release POWER MODE. The instrument returns to Gas Select Mode.
11. See “Exiting Gas Select Mode (START MEASURE)” on page 199 to return to Measuring
Mode.
Saving the Alarm Points (SAVE ALARM-P)
Performing a
SAVE ALARM-P
operation saves the current alarm setpoints.
Performing a
DEFAULT ALM-P
operation in the
ALARM SETTING
User Mode menu item
sets the instrument’s alarm points to those saved during the
SAVE ALARM-P
operation (if
performed).
Performing a
SAVE ALARM-P
operation has no effect on an
M.DEFAULT
in
Maintenance
Mode. An
M.DEFAULT
operation returns all instrument settings to the RKI default regardless of
if a
SAVE ALARM-P
operation was performed.
1. While in Gas Select Mode, use AIR to place the cursor next to
SAVE ALARM-P
.
2. Press and release POWER MODE. The display alternate between the following screens.
3. Press and release POWER MODE to save the current alarm point settings as the default.
4. The instrument returns to Gas Select Mode.
5. See “Exiting Gas Select Mode (START MEASURE)” on page 199 to return to Measuring
Mode.
MAINTE
7:49
GAS COMBO
SAVE ALARM--P
DISP MAX SPAN
GAS SELECT MODE
MAINTE
7:49
ARE YOU SURE ?
YES : MODE
SAVE ALARM-P
MAINTE
7:49
ARE YOU SURE ?
NO : AIR
SAVE ALARM-P