GX-2003 Operator’s Manual
Tips for Using Setup Mode • 59
Tips for Using Setup Mode
• When in the main menu, the cursor (>) will flash in front of a menu
item indicating that the item is selected.
• Use (SHIFT)
▼
to scroll down through the main menu and
submenus, and to lower values or change the setting in a specific
option.
• Use
▲
AIR to scroll up through the main menu and submenus, and
to raise values or change the setting in a specific option.
• Use the POWER ENTER button to enter a selected menu item, and
to enter and save settings during programming.
• An adjustable parameter that is flashing can be adjusted with the
▲
AIR and (SHIFT)
▼
buttons.
• Press the DISPLAY (ADJ) button while entering or updating
parameters to exit the screen without saving any changes.
Using Setup Mode
WARNING: The GX-2003 is not in operation as a gas detector while
in Setup Mode. Although it will respond to gas in parts
of AUTO CALIBRATION and SINGLE CALIBRATION,
there are no alarm indications.
1. Take the GX-2003 to a non-hazardous location, and turn it off if it
is on.
2. Press and hold down the
▲
AIR and (SHIFT)
▼
buttons, then press
and hold the POWER/ENTER button. When you hear a beep,
release the buttons. If the unit prompts you for the password, enter
it by using the
▲
AIR and (SHIFT)
▼
buttons to select each
password number and then pressing and releasing POWER
ENTER to enter it and move on to the next one. The main menu
displays. It displays four menu items at a time.
3. Use the
▲
AIR or (SHIFT)
▼
button to move the cursor up and
down the menu items and to view additional menu items.
>ALARM POINTS
LUNCH BREAK
ALARM LATCHING
ALARM SILENCE
21:09