10
•
Chapter 2: Description
65-2646RKS-SW M2A Transmitter Operator’s Manual
Housing Cap & Cap Gasket
The housing cap screws onto the detector housing. It retains the plug-in sensor and protects it
from damage. The housing cap also includes a flame arrestor which contains any sparks that may
occur within the detector and a flame arrestor guard which protects the flame arrestor from
damage. Unscrew the detector cap to access the plug-in sensor for maintenance or replacement. A
cap gasket seals the interface between the housing and cap. A flame arrestor guard is permanently
bonded to the cap.
Charcoal Filter and Rubber Boot
The disc-shaped charcoal filter is secured to the face of the CO sensor with a rubber boot. The
charcoal filter prevents interference gases (hydrogen sulfide [H
2
S] and certain hydrocarbons)
from producing false CO readings.
Plug-In CO Sensor
The sensor is secured within the detector assembly by the housing cap. Through a series of
chemical and electrical reactions, the sensor produces an electrical output that corresponds to the
detection range of the transmitter.
Magnetic Wand
The magnetic wand is a short plastic rod with a magnet in one end. It is used to actuate the
magnetic control switches on the control PCB while the junction box cover is still installed so that
non-intrusive calibration can be performed.