background image

 

65-2301RKSS Toxic Detector • 11

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended action for 
problems you may encounter with the toxic detector.

NOTE:

This troubleshooting guide describes detector problems only. See the controller operator’s 
manual for problems you may encounter with the controller.

Fail Condition

Symptoms

The controller indicates a fail condition.

Probable causes

The detector wiring is disconnected or misconnected.

The plug-in sensor is not properly plugged into the socket in the detector housing body.

The detector’s zero reading is low enough to cause a fail condition.

The detector is malfunctioning.

Recommended action

Verify that the detector wiring is correct and secure.

Confirm that the plug-in sensor is installed properly.

Perform a zero (fresh air) adjustment. A full calibration is recommended.

If the fail condition continues, replace the plug-in sensor as described later in this section.

If the fail condition continues, contact RKI for further instruction.

Slow or No Response/Difficult or Unable to Calibrate

Symptoms

Unable to accurately set the zero or response reading during the calibration procedure.

The detector requires frequent calibration.

NOTE:

Under “normal” circumstances, the detector requires calibration once every three months. 
Some applications may require a more frequent calibration schedule. See “Calibration 
Frequency” on page 13 f
or a discussion of the calibration frequency.

Probable causes

The calibration cylinder is low, out-dated, or defective.

The incorrect calibration cup or regulator is being used.

The plug-in sensor face is blocked with dirt or some other particulate contamination.

The detector is malfunctioning.

Recommended action

1.

Verify that the calibration cylinder contains an adequate supply of a fresh test sample.

2.

Confirm that you are using the correct calibration cup and regulator for your detector type. See 
Table 2 on page 14 for a list of the required calibration cups and regulators.

3.

Check the plug-in sensor face and remove any particulate contamination if necessary.

Summary of Contents for 65-2301RKSS Series

Page 1: ...www rkiinstruments com 65 2301RKSS Toxic Detector Operator s Manual Part Number 71 0273RK Revision A Released 7 17 19...

Page 2: ...alibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type o...

Page 3: ...t costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRES...

Page 4: ...the Toxic Detector to a Controller 8 Startup 10 Introducing Incoming Power 10 Setting the Zero Fresh Air Signal 10 Maintenance 10 Preventive Maintenance 10 Troubleshooting 11 Replacing Components of...

Page 5: ...2301RKSS toxic detectors WARNING When using the 65 2301RKSS you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65 2301RKSS and to minimize the ris...

Page 6: ...65 2300RK Toxic Detector The 65 2300RK toxic detector consists of the detector housing body detector housing cap splashguard cap gasket and the plug in sensor Figure 2 65 2300RK Toxic Detector Compone...

Page 7: ...as Through a series of chemical and electrical reactions the sensor produces an electrical output that corresponds to the detection range of the sensor Junction Box The stainless steel corrosion resis...

Page 8: ...g the Toxic Detector to a Controller WARNING Always verify that power to the controller is OFF before you make wiring connections 1 Turn off the controller 2 Turn off or unplug power to the controller...

Page 9: ...rough the same conduit hub The power cable may disrupt the transmission of the detector signal to the controller 8 Connect the wires to the applicable controller terminal strip See the controller oper...

Page 10: ...ccurate zero setting See Calibration on page 14 for instructions to introduce zero air to the detector 1 Verify that the detector is in a fresh air environment environment known to be free of toxic an...

Page 11: ...r in this section If the fail condition continues contact RKI for further instruction Slow or No Response Difficult or Unable to Calibrate Symptoms Unable to accurately set the zero or response readin...

Page 12: ...move the toxic sensor 5 Remove the black plug from the top of the replacement sensor 6 Carefully plug the replacement sensor into the connector that is located in the detector housing body WARNING You...

Page 13: ...llow the replacement detector to warm up for 5 minutes before you continue with the next step 11 Calibrate the replacement detector as described in Calibration on page 14 Calibration Frequency Althoug...

Page 14: ...ke sure to connect the sample tubing to the calibration cup port that results in the specified flow direction for the 81 1138RK calibration cup Preparing for Calibration 1 Unscrew the splashguard from...

Page 15: ...nder and toxic gas cylinder If necessary to fit the calibration toxic gas cylinder change the regulator Setting the Response Reading Span 1 Screw the regulator into the calibration gas cylinder 2 Foll...

Page 16: ...300RK SO2 Replacement detector assembly SO2 includes plug in sensor 71 0273RK 65 2301RKSS Operator s Manual this document 81 0076RK Zero air calibration cylinder 17 liter 81 0076RK 01 Zero air calibra...

Page 17: ...on cup and a 34 liter 10 ppm NH3 in nitrogen aluminum calibration cylinder 81 F603RK Calibration kit includes regulator calibration cup and 58 liter 2 ppm CL2 in nitrogen aluminum calibration cylinder...

Page 18: ...18 65 2301RKSS Toxic Detector ESM K01D CL2 10 ESM 01 plug in sensor 0 10 0 ppm chlorine diffusion type only Table 3 Parts List Part Number Description...

Reviews: