
61-1004RK/61-0191RK Combustible Gas Detector • 11
4.
Disconnect the detector leads from the terminal block inside the junction box. Note
the position of the color-coded leads as you remove them.
5.
Unscrew the detector from the junction box.
6.
Guide the detector leads of the replacement detector through the bottom conduit hub
of the junction box, then screw the mounting threads of the detector into the conduit
hub.
7.
Connect the detector leads to the terminal block in the same position as the leads you
removed in step 4.
8.
Secure the junction box cover to the junction box.
9.
Turn on power to the controller.
10. Turn on the controller.
11. Calibrate the replacement detector as described in“Calibration” on page 11.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 6 months is adequate for most infrared CO
2
detector
applications. Unless experience in a particular application dictates otherwise, RKI
Instruments, Inc. recommends a calibration frequency of every 6 months for the infrared
CO
2
detector.
If an application is not very demanding, for example detection in a clean, temperature
controlled environment where CO
2
is not normally present, and calibration adjustments
are minimal at calibration, then a calibration frequency of every 9 to 12 months is
adequate.
If an application is very demanding, for example if the environment is not well controlled,
then more frequent calibration than every 6 months may be necessary.
Calibration
This section describes how to calibrate the CO
2
detector. It includes procedures to
assemble the calibration kit, set the zero reading, set the response reading and return to
normal operation.
WARNING: The controller is not an active gas monitoring device during the calibration
procedure.
The following procedure assumes the use of a calibration kit which includes a 100%
nitrogen calibration gas cylinder for setting the zero, a CO
2
calibration gas cylinder for
setting the response reading, a 0.5 LPM fixed flow regulator with an on/off knob, a
calibration cup for the detector, and a short piece of sample tubing to connect the
regulator to the calibration cup.
Assembling the Calibration Kit
1.
Screw the calibration cup onto the detector.
2.
Use the sample tubing to connect the regulator to the calibration cup.
3.
Place the controller into its calibration program or disable external alarms.