82 • Appendix A: Maintenance Mode
SC-04
Turning the Zero Suppression On/Off (ZERO.SUP)
The
ZERO.SUP
setting is not intended for field adjustment. The default setting is
ON
.
Turning the Zero Follower On/Off (ZERO.FLW)
The
ZERO.FLW
setting is not intended for field adjustment. The default setting is
ON
.
User Mode Zero Suppression (ZSUP.DSP)
ON
: Zero suppression item appears in User Mode.
OFF
(factory setting): Zero suppression item does not appear in User Mode. The zero
suppression item is always available in Maintenance Mode.
It is not normally necessary to have the zero suppression item appear in User Mode. Contact
RKI Instruments before turning this setting on.
User Mode Zero Follower (ZFLW.DSP)
ON
: Zero follower item appears in User Mode.
OFF
(factory setting): Zero follower item does not appear in User Mode. The zero follower
item is always available in Maintenance Mode.
It is not normally necessary to have the zero follower item appear in User Mode. Contact RKI
Instruments before turning this setting on.
Cylinder Setting (CYL.DISP)
ON
:
CYL SEL
item appears in User and Maintenance Modes’
GAS CAL
item.
OFF
(factory setting):
CYL SEL
item does not appear in User and Maintenance Modes’
GAS CAL
item.
Sensor
Zero Suppression Value
Cl
2
0.10 ppm
HCN
0.5 ppm
NH
3
4 ppm
NO
2
0.30 ppm
PH
3
0.02 ppm
SO
2
0.20 ppm