Resin Pressure Sensor 

CZ-200P

 

Instruction Manual 

IM100CZ08-E9

 

Thank you for purchasing this RKC product. In order to achieve maximum 
performance and ensure proper operation of your new instrument, carefully 
read all the instructions in this manual. Please place the manual in a 
convenient location for easy reference. 

 

 

CONTENTS 

1. OUTLINE ···············································································································   1 
2. PRODUCT CHECK ·······························································································   2 
3. HANDLING OF THE PRESSURE SENSOR ·························································   2 

3.1 Caution in Mounting Pressure Sensors······························································   2 
3.2 Caution in Removing Pressure Sensors ····························································   4 
3.3 Cautions during Extruder Cooling Down ····························································   4 
3.4 Caution in Handling of Built-in Thermocouple ····················································   4 
3.5 Dimensions ········································································································   4 
3.6 Pressure Sensor Installation ··············································································   5 

4. WIRING ·················································································································   5 

4.1 Wiring Precaution ·······························································································   5 
4.2 Wiring Method ····································································································   6 

5. ADJUSTMENT ·······································································································   7 
6. CORRECTION ·······································································································   8 

6.1 Correcting Indication Error due to the Operating Temperature ··························   8 
6.2 Correcting Indication Error due to the Connection Cable Length ······················   8 
6.3 Correcting Indication Error when the Cable of Another Company is Used ············   8 
6.4 Correcting Indication Error due to the Safety Barrier ·········································   8 

7. TROUBLESHOOTING ···························································································   9 
8. SPECIFICATIONS ·································································································   9 
9. EXPLANATION OF EACH TERMS ·····································································   10 

 

 

SYMBOLS 

This mark indicates precautions that must be taken 
if there is danger of electric shock, fire, etc., which 
could result in loss of life or injury. 

 

This mark indicates that if these precautions and 
operating procedures are not taken, damage to the 
instrument may result. 

 

This mark indicates that all precautions should be 
taken for safe usage. 

 

This mark indicates important information on 
installation, handling and operating procedures. 

 

This mark indicates supplemental information on 
installation, handling and operating procedures. 

 

This mark indicates where additional information 
may be located. 

 

 

 

 

z

 An external protection device must be installed if failure 

of this instrument could result in damage to the 
instrument, equipment or injury to personnel. 

z

 All wiring must be completed before power is turned on 

to prevent electric shock, fire or damage to instrument 
and equipment. 

z

 This instrument must be used in accordance with the 

specifications to prevent fire or damage to instrument 
and equipment.   

z

 This instrument is not intended for use in locations 

subject to flammable or explosive gases. 

z

 Do not touch high-voltage connections such as power 

supply terminals, etc. to avoid electric shock. 

z

 RKC is not responsible if this instrument is repaired, 

modified or disassembled by other than 
factory-approved personnel. Malfunction can occur and 
warranty is void under these conditions. 

 
 
 

z

 This product is intended for use with industrial machines, test 

and measuring equipment. (It is not designed for use with 
medical equipment and nuclear energy.) 

 

 

z

 This is a Class A instrument. In a domestic environment, this 

instrument may cause radio interference, in which case the user 
may be required to take additional measures. 

z

 All precautions described in this manual should be taken to avoid 

damage to the instrument or equipment. 

z

 All wiring must be in accordance with local codes and regulations.   

z

 Tighten each terminal screw to the specified torque found in the 

manual to avoid electric shock, fire or malfunction. 

z

 Remove resin adhered to the instrument with dry cloth before the 

resin hardens. Take caution to avoid being burned. 

z

 Tools such as wire wheels or abrasive cloths should never be 

used to clean the process diaphragm. 

z

 Dropping the instrument may cause the sensor to break or 

malfunction. 

z

 To protect built-in precision components from damage, handle and 

install carefully. Take care to avoid damage to the diaphragm. 

z

 Do not wipe or rub the nameplate on the outer case of the resin 

pressure sensor with a cloth moistened with an organic solvent, or 
a glove. If so, the printed section may be erased. 

z

 This product uses stainless steel, aluminum, and fluororesin (O-rings 

and leadwire covering materials). When disposing of each part used 
for this product, always follows the procedure for disposing of industrial 
wastes stipulated by the respective local community. 

z

 If the output from the resin pressure sensor is found to be 

abnormal during the operation, immediately stop the operation and 
inspect for any distortion or damage of the diaphragm. If the 
diaphragm is damaged, the pressure of the measured media 
(e.g.resin) is applied to the inside of the resin pressure sensor. 
Continued use under such conditions may result in the damage of 
the screw which fixes the folder in the housing, and the folder in 
the housing may come off in the worst case. 

z

 Do not use the pressure sensor in stead of a blind bolt. 

 

NOTICE 

z

 This manual assumes that the reader has a fundamental knowledge 

of the principles of electricity, process control, computer technology 
and communications. 

z

 The figures, diagrams and numeric values used in this manual are 

only for purpose of illustration.   

z

 RKC is not responsible for any damage or injury that is caused as a 

result of using this instrument, instrument failure or indirect damage. 

z

 RKC is not responsible for any damage and/or injury resulting 

from the use of instruments made by imitating this instrument. 

z

 Periodic maintenance is required for safe and proper operation of 

this instrument. Some components have a limited service life, or 
characteristics that change over time. 

z

 Every effort has been made to ensure accuracy of all information 

contained herein. RKC makes no warranty expressed or implied, 
with respect to the accuracy of the information. The information in 
this manual is subject to change without prior notice. 

z

 No portion of this document may be reprinted, modified, copied, 

transmitted, digitized, stored, processed or retrieved through any 
mechanical, electronic, optical or other means without prior written 
approval from RKC. 

 

1. OUTLINE 

„

 

Great features 

 As its lead pipe and pressure sensing blocks have a triple 

construction output indication changes caused by external transient 
temperature variations are extremely small. In addition, the 
attachment of a lead pipe cover (optional) further restricts small 
output indication changes. 

  As the sensor uses the push rod method, resin contamination is 

prevented for the sensor damage. 

  High accuracy of the measured pressure can be achieved when the 

instrument is used in conjunction with an output converter having 
linearizing function (PCT-300) or a pressure indicator (PG500 or 
REX-PG410). However, excluding the sensor having pressure 
range (70 MPa or more) and a HASTELLOY C diaphragm. 
 Overall 

accuracy: 

±

0.5 % of full scale or less 

 

 

PCT-300 

Pressure sensor
CZ-200P 

0 to 10 V 
0 to 10 mV 
1 to 5 V 
4 to 20 mA 

Recorder

Pressure

controller 

4-core sielded cable 

Indicator

Alarm 

PG500 or 

REX-PG410 

or 

0 to 10 V 
0 to 10 mV 
0 to 2 V 
0 to 5 V 
1 to 5 V 
4 to 20 mA 
0 to 20 mA 

 

 

 

 

WARNING

 

CAUTION

 

!

 

!

 

WARNING 

CAUTION 

Summary of Contents for CZ-200P

Page 1: ...rned z Tools such as wire wheels or abrasive cloths should never be used to clean the process diaphragm z Dropping the instrument may cause the sensor to break or malfunction z To protect built in precision components from damage handle and install carefully Take care to avoid damage to the diaphragm z Do not wipe or rub the nameplate on the outer case of the resin pressure sensor with a cloth moi...

Page 2: ...C diaphragm specification is not available 4 The SUPRON diaphragm specification is not available to the range of 5 MPa or less and 150 MPa z Accessories Instruction Manual IM100CZ08 E9 Copper Packing Included in only the Loose nut type thickness t 2 mm If any of the products are missing damaged or if your manual is incomplete please contact RKC sales office or the agent Temperature sensor cable So...

Page 3: ...the exposed lead pipe is further extended In this case take measures of 1 The effect of heat flow lessens as the installing direction of the sensor changes from the slanting direction Fig 1 B to the horizontal direction Fig 1 C in this order In this case take measures of 1 and 2 if necessary by checking the sensor outer cylinder surface temperature To the relevant manufacturer It is recommended th...

Page 4: ...rush or bend the thermocouple protecting flexible tube extended from the outer cylinder or the thermocouple plug If so the thermocouple may be damaged to disable temperature measurement or a temperature error may result The temperature sensing block is located about 2 mm within the diaphragm and as a result the measured temperature usually becomes lower than the actual temperature depending on the...

Page 5: ...pe cover Optional The outer tube and the cable section of the loose nut type and the unified thread type with directly connected cable are the same as the above figure The mounting hole dimensions are the same as those for the standard type PF3 8 JIS B grade Diaphragm Unit mm External dimensions Standard dimensions H A H B H C H D L 120 150 180 210 L1 8 8 8 8 L2 6 6 6 6 L3 60 90 120 150 Weight App...

Page 6: ...21 Colors Terminal No Input terminals Wiring with REX PG410 pressure indicator Pressure sensor cable Pressure sensor CZ 200P SHD E EXC A EXC C SIG D SIG B REX PG410 Shield Red Brown Blue Black 6 7 8 9 10 Colors Terminal No Input terminals For details on wiring with PCT 300 PG500 or REX PG410 refer to the relevant instruction manual Wiring example of built in temperature sensor type Optional Pressu...

Page 7: ...setter Monitor plug Zero adjuster Power lamp PCT 300 PWR ZERO mV V MONI 1 Check the rated output mV V described on the nameplated of the CZ 200P This output should be corrected when the cable is extended and then set that value on the rotary switch which is gain setter of the PCT 300 2 The pressure reading zero point is adjusted by the zero adjuster in PCT 300 Perform this zero adjustment after th...

Page 8: ...al length L 10 m Rated output of the resin pressure sensor e0 1 500 e1 1 500 1 1 96 10 4 10 5 e1 1 499 Set 1 499 to the gain setting portion of the converter 6 3 Correcting Indication Error when the Cable of Another Company is Used The nominal cross sectional area of our cable conductor is 0 5 mm 2 If using a cable from another company the resin pressure sensor s indication value can be corrected ...

Page 9: ...ation or damage refer to the Main causes on page 9 Main causes These causes may arise independently or in mutual relations Generative cause in operation Overpressure Load pressure exceeding its limit applied Irregular thread engagement The metal pressure sensing surface mechanically scratched or chipped off Metal fatigue The metal sensing surface fatigued by the application of changing or repeatin...

Page 10: ... diaphragm specification type As to diaphragm temperature 0 15 0 2 of span 10 C Mechanical characteristics Allowable over pressure 120 of span 0 5 MPa 1000 of span 1 MPa 500 of span Limited over pressure 150 of span 0 5 MPa 2000 of span 1 MPa 1000 of span Fixing screw section material SUS630 Lead pipe cover material SUS304 Fixed nut type only Diaphragm material SUS630 HASTELLOY C SUPRON Recommende...