Rixen TK-6200 Instruction Manual Download Page 11

 

**Example** 

Set mode:         
1. Temperature unit=℃ 
2. No use 
3.T1 no ERROR 
4. T2 no ERROR 
5. Power74~50% 

Set    mode    MOD 

Bit 15 

Bit 14 

Bit 13 

Bit 12 

Bit 11 

Bit 10 

Bit 09 

Bit 08 

 

Bit 07 

Bit 06 

Bit 05 

Bit 04 

Bit 03 

Bit 02 

Bit 01~ Bit 00 

℉=1/℃=0 

T2 Error=1 

T1 Error=1 

Power 

☆Power Meter is divided into 4 parts    11=100~75%            10=74~50% 

                                                       

            01=49~25%              00=24~0% 

 

 

-10- 

 

Address 

Data name 

Data size 

0000h 

T1 Real-Time Data(Hold) 

4-Byte 

0001h 

T1 Max 

4-Byte 

0002h 

T1 Min 

4-Byte 

0003h 

T1 Avg 

4-Byte 

0004h 

T2 Real-Time Data(Hold) 

4-Byte 

0005h 

T2 Max 

4-Byte 

0006h 

T2 Min 

4-Byte 

0007h 

T2 Avg 

4-Byte 

0008h 

First 2-byte = T1-Hi.A 
Last 2-byte = T1-Lo.A 

4-Byte 

0009h 

First 2-byte = T2-Hi.A 

Last 2-byte = T2-Lo.A 

4-Byte 

000Ah 

Set mode = MOD 

2-Byte 

Summary of Contents for TK-6200

Page 1: ...CONTENTS TK 6200 specifications 1 Instrument descriptions 2 Display descriptions 3 Key descriptions 4 Instructions 5 RS 232 transfer protocol 8 Precautions 11...

Page 2: ...protocol provided Power source 9 V Battery or AC Adapter Dimensions Weight 150 x 75 x 28 mm Approx 320g battery included Input connection Mini connector waterproof socket IP66 Standard Applications TK...

Page 3: ...n LCD Display 7 Battery cover 2 Power switch 8 AC Adaptor Input 3 Setting and switching 9 USB Output port 4 Function key 10 RS 232 Output connector 5 Setting 11 T1 Probe input connector 6 Model No and...

Page 4: ...ay either one testing probe is Abnormal 4 If is appears when turning on the instrument please let go off all keys and tune Instrument on again 3 Display Descriptions Battery Power Symbol Manual shut d...

Page 5: ...e after data confirmed please press for over 2 seconds again for enter the next level Push key to increase the value Push key to reduce the value Push key to change digit Determine the value re hold k...

Page 6: ...emp data Area A T2 Hi A data Area B T2 Lo A data Area C T1 T2 Real time Temperature display T1 T2 Temp data Area A T1 data Area B T2 data Area C Warning alert When the temperature is higher than the H...

Page 7: ...system are changed Condition Display content T1 HOLD display Lock the T1 Temperature data T1 MAX display T1 Maximum Temperature data T1 MIN display T1 Minimum Temperature data T1 AVG display T1 Avera...

Page 8: ...m settings Lowest value 200 0 328 0 Deviation setting range 100 00 Start with 0 00 180 00 Start with 0 00 Span setting range 0 00 200 00 starts with 100 00 7 Display content Description baud display B...

Page 9: ...i Starting address Lo No of Byte Hi No of Byte Hi CRC Lo CRC Hi 02H 03H 00H 00H 00H 04H 44H 3AH Response Data Frame Ex Response value 4 Byte Response Data 80003 Slave address Function Byte count Data...

Page 10: ...te Slave address Function Byte count Data 1 Hi Data 1 Middle Data 1 Lo Data 1 Last 02H 03H 0AH 03H E8H FCH 18H Data 1 Hi and Data 1 Middle is value of T1 Hi A 0x03E8 1000 Data 1 Lo and Data 1 Last is...

Page 11: ...1 T1 Error 1 Power Power Meter is divided into 4 parts 11 100 75 10 74 50 01 49 25 00 24 0 10 Address Data name Data size 0000h T1 Real Time Data Hold 4 Byte 0001h T1 Max 4 Byte 0002h T1 Min 4 Byte 00...

Page 12: ...ed 3 Please keep the temperature stabled when measuring the temperature of an object to obtain a more accurate result 4 When the instrument shows power shortage warning please immediately replace the...