3.2.1.10 ENERGY DIGIT RESET COUNT (ROLLOVER COUNT)
This screen enables the user for setting maximum energy count after which energy will
rollover to zero depending on the setting of Wh, kWh & Mwh in Energy resolution option.
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
CT SECONDARY
5 AMPERE
1 AMPERE
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
ENERGY DIGIT RESET COUNT
7 DIGITS
9 DIGITS
11 DIGITS
8 DIGITS
10 DIGITS
12 DIGITS
13 DIGITS
14 DIGITS
OK
BACK
If Energy
resolution
is in WATT then rollover count can be from 7 to 14 DIGITS.
If Energy
resolution
is in KILO-WATT then rollover count can be from 7 to 12 DIGITS.
If Energy
resolution
is in MEGA-WATT then rollover count can be from 7 to 9 DIGITS.
Touching radio button in front of particular option will
select that option.
Touch on “ key” will conrm the setting.
OK
OK
Touching the “ key” will keep the old selected setting
BACK
and will return to previous menu.
Note :-
1) Default value of energy digit reset count is set to “14” i.e if energy crosses the
14 digit count it will rollover to zero.
2) If Energy Resolution is set to kW & energy digit reset count is set to 12, Energy screen
on display will show “-------” i.e energy overow when energy crosses the 11 digit count.
3) If Energy Resolution is set to MW & energy digit reset count is set to 9, Energy screen
on display will show “-------” i.e energy overow when energy crosses the 8 digit count.
3.2.1.9 Energy Update Rate
ENERGY UPDATE RATE
ENTER ENERGY RATE IN MIN
15