background image

Page 5

DLS350 User Guide

Basic Testing

The DLS350 is calibrated to have an output power level of -22dBm. Variations in power level 

between -22dBm and -25dBm may be normal depending on the quality and age of the test jumper, the DLS350 output 

port, and other factors.

Always ensure the DLS350 is transmitting an acceptable power level before performing an insertion loss test. Simply insert 

the test jumper (plugged into the DLS350) into a companion power meter set to the dBm mode. The power meter will 

indicate the measured output power of the LED. If this reading is not near -22dbm first inspect the DLS350 ferrule with an 

Inspection Scope. If it passes according to IEC try using a different  test jumper. If none of these resolve your issue please 

contact Tech support -see the final page of this document for contact information.

Transmitting Light

The DLS350 transmits either the 850nm or 1300nm wavelength on Multimode fiber. Be sure to use a test jumper to mate 

the DLS350 to the fiber under test. this allows you to isolate the DLS350 from any dirt on the fiber being tested and 

allows you to set reference when doing relative power tests.

Test Jumper

Mate Connectors 

Signal Propagates Down 

the Fiber

Caution: Invisible LED Radiation

Please note that 850nm and 1300nm wavelengths are not visible to the human eye. Do not look directly into the output 

port of the DLS350 or directly into any fiber connector that may be live.

Since the LED is invisible to the eye, the eye’s natural blink reflex is suppressed. This can cause damage to the retina.

FDA 21 CFR 1040.10 and 1040.11

IEC 60825-1: 2007-03

Class 1 LED Product

Output Power

Test Jumper Plugged 

Into Power Meter

Power Meter Displays 

Output Power of LED

Summary of Contents for Miller DLS350

Page 1: ...Page 1 DLS350 User Guide DLS350 Dual LED Source User Guide ...

Page 2: ...Page 2 DLS350 User Guide Contents Introduction 3 Device Overview 3 Important Safety Information 4 Basic Testing 5 Using 2kHz 6 Maintenance 6 Certifications and Contact Information 7 ...

Page 3: ... the DLS350 ON and OFF The unit will turn off automatically after 15 minutes by default To disable this timeout hold the power button for at least 5 seconds when turning ON 2kHz Button Toggles 2kHz tone on the currently selected wavelength This is usefull for detecting which fiber is being tested at the other end however the modulation reduces output power by Wavelength Button cycles through eithe...

Page 4: ...rd non reflective surfaces to prevent reflection of laser Electric Shock Hazard Pay attention to proper battery polarity Do not mix battery types or manufacturers Do not open the unit with the exception of the battery compartment door Use this unit only for its intended purpose as outlined in this document Damage to Item Hazard Do not leave item in direct sunlight or near heat sources submerge in ...

Page 5: ...Tech support see the final page of this document for contact information Transmitting Light The DLS350 transmits either the 850nm or 1300nm wavelength on Multimode fiber Be sure to use a test jumper to mate the DLS350 to the fiber under test this allows you to isolate the DLS350 from any dirt on the fiber being tested and allows you to set reference when doing relative power tests Test Jumper Mate...

Page 6: ...pers must be inspected and cleaned before plugging into the DLS350 unit If soiled or damaged connectors are inserted they can cause damage to the DLS350 output port and the unit may need to be repaired The test jumper has physical contact with the ferrule inside the DLS350 To clean the ferrule inside the DLS350 Unplug the test jumper Unscrew the adapter until it spins freely then pull Inspect the ...

Page 7: ...e recalibration within 2 years of the date of purchase To start the calibration process simply call or email technical support Calibration Certificates The DLS350 comes with a 2 year warranty for any manufacturer defects or damage due to reasonable use To start the RMA process simply email or call technical support Warranty LED SOURCE Wavelength 850nm 1300nm Output Power 22 00 dBm Output Stability...

Page 8: ...s are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equ...