![background image](http://html1.mh-extra.com/html/rion/rionet-hb-l1/rionet-hb-l1_instruction-manual_1463471002.webp)
Volume switching
The selector button gives access to four different volume level
settings *. Using the selector button in the way shown below
lets you switch between the settings. When power to the hear-
ing aid is turned on (the battery holder is closed), the volume
setting will be "Level 1". Depending on the environment (noisy
location, quiet location, etc.) you may want to change the setting.
When you operate the button, a beep sound pattern indicates
the change.
* These have been optimized for your individual requirements.
No. 51970 07-06
you switch between the settings. When power to the hearing
aid is turned on (the battery holder is closed), the setting will be
"Memory 1". Depending on the environment (noisy location, quiet
location, etc.) you may want to change the setting. When you op-
erate the button, a beep sound pattern indicates the change.
* These have been optimized for your individual requirements.
There may be from 1 to 4 different memory settings.
120
0dB=20µPa
1600 2500
110
100
90
80
70
100
200
500
1000
2000
5000
10000
Frequency (Hz)
O
ut
pu
tS
P
L
(d
B
)
Input SPL 60dB
130
0dB=20µPa
1600 2500
120
110
100
90
80
100
200
500
1000
2000
5000
10000
Frequency (Hz)
O
ut
pu
tS
P
L
(d
B
)
Input SPL 90dB
60
1600 2500
50
40
30
20
10
100
200
500
1000
2000
5000
10000
Input SPL 50dB
(0dB=20µPa)
Frequency (Hz)
A
co
us
tic
G
ai
n
(d
B
)
120
0dB=20µPa
1600 2500
110
100
90
80
70
100
200
500
1000
2000
5000
10000
Frequency (Hz)
O
ut
pu
tS
P
L
(d
B
)
Input SPL 60dB
60
1600 2500
50
40
30
20
10
100
200
500
1000
2000
5000
10000
Input SPL 60dB
(0dB=20µPa)
Frequency (Hz)
A
co
us
tic
G
ai
n
(d
B
)
Full-on gain curve
Frequency response curve
OSPL90 curve
Full-on acoustic gain
frequency response curve
Basic frequency response
curve
Note:
(1) Production number (serial
number) is indicated at
the upper rear side of the
unit.
(2) Manufacturer's name is
imprinted at the lower rear
side of the unit.
TECHNICAL DATA
(
According to ANSI standard S3.22 2003
)
Maximum OSPL90
113 dB (660 Hz)
HFA-OSPL90
107 dB
HFA-full-on Acoustic Gain 38 dB
Reference Test Gain
30 dB
Frequency Range
200 Hz to 5000 Hz
Equivalent Input Noise Level
31 dB
Total Harmonic Distortion 500 Hz: 1% 800 Hz: 1% 1600 Hz: 1%
AGC
(Attack / Recovery Time)
Input AGC
Attack:
3 msec or 400 msec
Recovery: 100 msec or 600 msec or
15 sec
Output AGC
Attack:
3 msec
Recovery: 50 msec
Battery Type/Supply Voltage 312/1.4 V
Battery Current/Battery Life 0.90 mA/160 hours
Dimensions/Weight
3.82 × 2.03 × 0.80 cm /
1.9 g (excluding battery)
(Typical value)
TECHNICAL DATA
(
)
Reference Test Frequency 1600 Hz
OSPL
90
113 dB
500 Hz
116 dB
Peak
117 dB
Full-on Acoustic Gain
43 dB
Equivalent Input Noise Level 30 dB
Total Harmonic Distortion 500 Hz: 1% 800 Hz: 1% 1600 Hz: 1%
AGC
(Attack / Recovery Time)
Input AGC
Attack:
3 msec or 400 msec
Recovery: 100 msec or 600 msec or
15 sec
Output AGC
Attack:
3 msec
Recovery: 50 msec
Battery Type/Supply Voltage PR41 (312)/1.4 V
Battery Current/Battery Life 0.85 mA/170 hours
Dimensions/Weight
3.82 × 2.03 × 0.80 cm /
1.9 g (excluding battery)
(Typical value)
According to IEC standard Pub. 60118-0-1983
Amendment 1-1994
Level
1
(
Medium*)
2 beeps
Beep sound
Power-on
Beep sound
Beep sound
Level
2
(High*)
Level
3
(
Maximum*)
Level
4
(
Low*)
3 beeps
4 beeps
Beep sound
1 beep
* These are the standard volume settings.
They can be changed by your hearing aid professional.
Memory
1
2 beeps
Beep sound
Power-on
Beep sound
Beep sound
Memory
2
Memory
3
Memory
4
3 beeps
4 beeps
Beep sound
1 beep
Memory switching
The selector button gives access to four different memory set-
tings *. Using the selector button in the way shown below lets
Turn-on precaution
If the hearing aid is turned on again immediately after turning it off,
there may be no sound, but this is not a defect. In such a case,
turn the power off (open battery holder) and wait 10 seconds.
Then turn the power on again (close battery holder).
Adjusting the Controls
This hearing aid is a digital type. Any adjustments must be done
at the dealer. For details, please contact the dealer.
130
0dB=20µPa
1600 2500
120
110
100
90
80
100
200
500
1000
2000
5000
10000
Frequency (Hz)
O
ut
pu
tS
P
L
(d
B
)
Input SPL 90dB
OSPL
90
curve