![Rion NC-74 Instruction Manual Download Page 30](http://html1.mh-extra.com/html/rion/nc-74/nc-74_instruction-manual_1463506030.webp)
18
Reference
Immunity to power-frequency magnetic fi elds and radiofrequency
electromagnetic fi elds
Sound pressure level deviation when placed under the infl uence of power-frequency mag-
netic fi elds and radiofrequency electromagnetic fi elds as specifi ed below:
within ±0.3 dB
- Electromagnetic fi eld strength up to 10 V/m rms (non-modulated), frequency range
26 MHz to 1 GHz, 900 Hz sinusoidal wave, 80% amplitude modulation
- AC magnetic fi eld strength up to 80 A/m rms, frequency 50 Hz and 60 Hz
The confi guration that produce minimum immunity (maximum susceptibility) to power-
frequency magnetic fi elds and radiofrequency electromagnetic fi elds:
power ON