![Rion NC-74 Instruction Manual Download Page 22](http://html1.mh-extra.com/html/rion/nc-74/nc-74_instruction-manual_1463506022.webp)
10
Measurement
7. Adjust the sound level meter so that its level reading is 94 dB (calibration value, see
Note below).
Note
Depending on sound level meter model and type, there may be
deviation of 0.1 to 0.5 dB, due to the infl uence of case refl ections
and the microphone diffraction effect.
For details on precise calibration values, please refer to the "Ref-
erence" section on page 11 or to the documentation of the sound
level meter.
8. When level adjustment is completed, set the power switch of the sound level meter
and the NC-74 to OFF.
9. Carefully remove the microphone of the sound level meter from the NC-74 (see
the “Important” on page 7).
Summary of Contents for NC-74
Page 2: ......
Page 6: ...iv ...
Page 10: ...viii ...
Page 12: ...x ...
Page 37: ......
Page 38: ...No 32263 14 08 ...