10.DETECTION PRINCIPLE 

 
 
This  gas  detector  applies  an  electrochemical  sensor.  The  sample  gas  is  electrolyzed  by  the 
electrolyzed  cell  added  with  specific  electric  potential  (bias  voltage)  and  detected  from  the 
electrolyzed current generated at that time. 
The electrochemical sensor is designed to keep the interface between electrode and electrolyte 
at a constant potential (Bias voltage) and is the method to electrolyze gas directly. Then as the 
gas has the bias voltage generating its own electrolization (Redox reaction) the bias voltage of 
sensor is determined by the redox potential. 
 
 
 
 
 
 
        WE 
                              REF                                                                                                            Output 
 
        CE 
 
 
 
 
 
 
 
 
 
 

  

P

o

t

e

n

t

i

o

s

t

a

t

 

c

i

r

c

u

i

t

 

 

 

WE 

: Working electrode 

CE 

: Counter electrode 

RE 

: Reference electrode 

Summary of Contents for GD-K71D

Page 1: ...PT2E 1057 OPERATING MANUAL FOR INTELLIGENT GAS DETECTOR GD K71D...

Page 2: ...operation 13 Installation place 13 Caution in the system engineering 14 Grounding 17 Caution at wiring construction 18 Applicable cable Power signal cable 18 Relay output signal cable 18 Specification...

Page 3: ...contract for regular check 33 Calibration method Calibration with gas etc 34 Replacement of sensor and parts 34 Treatment when operation is stopped or its installation place is moved Stopping the norm...

Page 4: ...4 digit LED indicator The GD K71D transmits gas concentration signals with 4 20mADC to upper system This detector provides 2 relay outputs for outer gas alarm Identification of each signal word Indic...

Page 5: ...inside of the panel and the power source as well Do not use fuse that is not specified or do not make a short circuit inside of the fuse holder Operation in gas Do not operate the instrument in an atm...

Page 6: ...he lower red bar the correct gas detection cannot be performed Adjust the flow rate Equip a dust filter to the outside of the detector if it is used in a dusty place The dust filter depends on a measu...

Page 7: ......

Page 8: ......

Page 9: ......

Page 10: ...bracket on the wall Put the detector unit on the wall mounting bracket Confirm that the detector unit is surely fixed to the wall mounting bracket after finishing the job If it is not surely fixed th...

Page 11: ......

Page 12: ......

Page 13: ......

Page 14: ...electronics parts Install them where to be stable not to crush or fall down Keep the system from the equipment which may give a high frequency in the surroundings and install it Do not put the system...

Page 15: ...ntact may be on and the care for it must be taken In such case use the standby battery or take an appropriate action in the receiver side Supply the following power to this unit Power voltage DC V Pow...

Page 16: ...hunder lightning or except it To protect detector from these cause be sure to make grounding In the lightning rod there is the circuit to remove the surge voltage to be a cause of damage from field in...

Page 17: ...elay is melted adhesively and the contacts can not work High voltage is put inside of detector unit and then electrical parts may be damaged As it is big noise the trouble action may be taken by the r...

Page 18: ......

Page 19: ...be used Bar terminal Model AI series Phoenix contact make Terminal lug tool Model CRIMPFOX UD 6 Phoenix contact make CAUTION Be sure to use the exclusive bar terminal When used with other make bar ter...

Page 20: ...Connection with indicator unit DCS PLC 4 wire type Trouble contact ALM contact ALM contact 4 20mA EC 583 EC 573A DC24V NOTE Terminals 4 5 are not in use 2 3 4 5 E GD K71D Detector E 1 2 3 4 5 6 E E GN...

Page 21: ......

Page 22: ...Arrange not to affect the outside even if the contact work Check that the dust filters are connected correctly or not For security purpose check that it is a designated rated fuse or not Basic perfor...

Page 23: ......

Page 24: ......

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ......

Page 29: ......

Page 30: ......

Page 31: ...heck the external output and function of outer equipment to be connected with external alarm contact Then judge whether power can be shut off or not When alarm contact is set on energized condition op...

Page 32: ...ALM1 Orange ALM2 Red There are two alarm levels When the reading goes to alarm preset level or exceeds it this turns to on 2 External output 4 20mA output Electric current in proportion to gas concent...

Page 33: ...alarm possibility except the case of gas detection It may respond to interference gas For interference gases contact the nearest agent or Riken Keiki It may be caused by drifting due to the change of...

Page 34: ...While this unit is working ball in flowmonitor shall be positioned within two red lines If it is below the lower part of red line adjust it with FLOW ADJ switch to Even though it cannot move up the fi...

Page 35: ...a dark place under the normal temperature and humidity away from direct sunlight In a place where gases solvents or vapors are not present CAUTION When move its installation place be sure to make gas...

Page 36: ...ring Connection ports of sensor unit and chassis unit 1 year 3 6 4 9 Lithium battery 3 1 10 Flowmonitor ass y 1 year 7 8 1 11 O ring Flowmonitor part 1 year 3 6 2 12 Dust filter Outside use 6 months...

Page 37: ...lter or deterioration of pump Counteraction at trouble alarm When any trouble was found on this gas detector contact our nearest agent or Riken Keiki If trouble should take place we try best to search...

Page 38: ...easing of sensor sensitivity The sensor replacement is required 4 PW TR light flashing System Trouble E 00 Probable cause and Recommended action Memory trouble inside of detector Contact our nearest a...

Page 39: ...e dust filter Is the pipe at gas inlet side or gas outlet side broken or clogged Repair or replace defective part Check whether detector unit is firmly connected 6 Fuse disconnection Symptom The detec...

Page 40: ...e used with this detector requires an electric potential even if it is not in use This is a function to supply electric potential to the sensor when it is not in use or power suspension PPM This is th...

Page 41: ...e function to suspend the running temporarily Alarm response time It is the time until the alarm is given when exposed to gas concentration of 1 6 times alarm preset level This does not involve the ti...

Page 42: ...The electrochemical sensor is designed to keep the interface between electrode and electrolyte at a constant potential Bias voltage and is the method to electrolyze gas directly Then as the gas has th...

Page 43: ...lighting Relay output Dry contact normally open for both 1st and 2nd alarm rated 0 25A at 125VAC 0 5A at 24VDC Resistive load Alarm delay time Within 60sec Time to get an alarm when exposed to gas con...

Page 44: ...perature 0 40 Humidity 30 70 R H standard Consult with us other humidity Structure Box type wall mounting type Outer dimensions 70 W X170 H X153 D mm Weight Approx 2 0kg 11 Product composition Standar...

Page 45: ...ne B2H6 0 0 3ppm 0 1 ppm M Dichlorosilane SiH2Cl2 HCl 0 15ppm 5 ppm M Disilane Si2H6 0 15ppm 5 ppm M B Fluorine F2 0 3ppm 1 ppm M Germane GeH4 0 0 8ppm 0 2 ppm M B Germanium Tetrachloride GeCl4 HCl 0...

Page 46: ...9ppm 3 ppm M Sulfur Tetrafluoride SF4 HF 0 9ppm 3 ppm M Tantalum Fluoride TaF5 HF 0 9ppm 3 ppm M Tin Tetrachloride SnCl4 HCl 0 15ppm 5 ppm M Trichlorosilane SiHCl3 HCl 0 15ppm 5 ppm M Tungsten Hexach...

Page 47: ......

Reviews: