6 Maintenance
6-2. Replacement parts
- 32 -
6-2. Replacement parts
<Replacement of Gas Sensor>
Our service engineers need to replace and adjust the sensor. Please contact RIKEN KEIKI.
NOTE
If adjustment to the standard gas concentration value fails even with the maximum sensitivity, it
indicates that the gas sensor has come to the end of its life. The gas sensor needs to be replaced.
After replacing the gas sensor, electrical adjustment and calibration using the standard gas are
necessary.
<Replacement of Regular Replacement Parts>
List of recommended regular replacement parts
Name
Maintenance
intervals
Replacement
intervals
Quantity
(pieces per
unit)
Remarks
Elbow
6 months
3 - 8 years
2
O-ring (flow regulator)
6 months
3 - 8 years
1
O-ring (sensor cap)
6 months
3 - 8 years
1
Tube
6 months
3 - 8 years
710 mm
Tube
6 months
3 - 8 years
50 mm
For the stable operation of the detector and safety, ask a qualified service engineer to take care of
replacement of the parts. Request the operation check from RIKEN KEIKI.
NOTE
The above replacement interval is recommendation only. The interval may change depending on the
conditions such as operating environment. These intervals do not mean the warranty periods either. The
result of the regular maintenance may determine when to replace the parts.