7 Maintenance
7-2. Regular maintenance mode
- 61 -
RM-6000
SE 3. GAS SEL
Select and set the type of gas of
measured gas and the full scale.
Press the
▲
or
▼
key to make a
selection, and then press the SET
key to confirm the selection.
This is not typically used by the
user.
SE 4. PEAKHOLD
Select either ON/OFF of holding
the maximum concentration value
when a gas is detected.
Press the
▲
or
▼
key to select
either ON/OFF, and then press
the SET key to confirm the
selection.
SE 5. ALM1 LCD
Select the LCD color (orange or
red) when the first alarm is
triggered.
Press the
▲
or
▼
key to select a
color, and then press the SET key
to confirm the selection.
SE 6. - - -
(No environmental setting is
available.)
Setting of Type of Gas
of Measured Gas and
Full Scale
=> P67