5 How to Operate
5-7. User mode
RM-6000
- 34 -
5-7. User mode
<<User Mode>> (The following is display example of GP-6001.)
WARNING
After the adjustment is completed, never fail to press the MODE key to return to the detection mode.
(If the indicator/alarm unit remains in the user mode, it automatically returns to the detection mode
in ten hours.)
Detection Mode
Press the MODE key for
three seconds.
User Mode
1-1. ZERO
Perform the zero
adjustment.
(In case of OX-6001 and
OX-6002, perform the span
adjustment.)
1-2. CONFIRM
Show the setting of the
typical menu.
1-3. PEAK
Displays the maximum
concentration value when a
gas is detected.
1-4. RM VER
Show the program version
of the main unit.
1-5. AMP VER
Show the program version
of the AMP unit.
1-6. ADDRESS
Show the communication
address of RS-485.
(Only on a model with an
option mounted)
1-7. 485 PTRN
Show the communication
setting of RS-485.
(Only on a model with an
option mounted)
Zero Adjustment
=> P36
Setting Display
=> P37
During maintenance
During inhibit
I