![Rigol RSA5032 Programming Manual Download Page 104](http://html1.mh-extra.com/html/rigol/rsa5032/rsa5032_programming-manual_1461068104.webp)
RIGOL
2-90
VSA Programming Guide
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
OFF|0
Return Format
The query returns 1 or 0.
Example
The following command enables the trigger holdoff function.
:TRIGger:SEQuence:HOLDoff:STATe ON or :TRIGger:SEQuence:HOLDoff:STATe 1
The following query returns 1.
:TRIGger:SEQuence:HOLDoff:STATe?
:TRIGger[:SEQuence]:LEVel
Syntax
:TRIGger[:SEQuence]:LEVel <ampl>
:TRIGger[:SEQuence]:LEVel?
Description
Sets the trigger level of the IF power trigger.
Queries the trigger level of the IF power trigger.
Parameter
Name
Type
Range
Default
<ampl>
Consecutive Real Number
-140 dBm to 30 dBm
0 dBm
Remarks
This command is only valid when IF power trigger is selected.
Return Format
The query returns the trigger level in scientific notation.
Example
The following command sets the trigger level of IF power to 10 dBm.
:TRIGger:SEQuence:LEVel 10
The following query returns 1.01.
:TRIGger:SEQuence:LEVel?
:TRIGger[:SEQuence]:SOURce
Syntax
:TRIGger[:SEQuence]:SOURce EXTernal1|EXTernal2|IMMediate|FMT|POWer
:TRIGger[:SEQuence]:SOURce?
Description
Sets the trigger source.
Queries the trigger source.
Parameter
Name
Type
Range
Default
--
Keyword