RIGOL
VSA Programming Guide
2-81
Return Format
The query returns the time value in scientific notation. The unit is s.
Example
The following command sets the time to 10 ms.
:TRIGger:SEQuence:ATRigger 0.01
The following query returns 1.000000000e-02.
:TRIGger:SEQuence:ATRigger?
:TRIGger[:SEQuence]:ATRigger:STATe
Syntax
:TRIGger[:SEQuence]:ATRigger:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:ATRigger:STATe?
Description
Enables or disables the auto trigger function.
Queries the setting status of auto trigger function.
Parameter
Name
Type
Range
Default
——
Bool
OFF|ON|0|1
OFF|0
Return Format
The query returns 1 or 0.
Example
The following command enables the auto trigger function.
:TRIGger:SEQuence:ATRigger:STATe ON or :TRIGger:SEQuence:ATRigger:STATe 1
The following query returns 1.
:TRIGger:SEQuence:ATRigger:STATe?
:TRIGger[:SEQuence]:EXTernal<n>:DELay
Syntax
:TRIGger[:SEQuence]:EXTernal<n>:DELay <time>
:TRIGger[:SEQuence]:EXTernal<n>:DELay?
Description
Sets the delay time for the external trigger.
Queries the delay time for the external trigger.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2
--
<time>
Consecutive Real Number
0 μs to 500 ms
1 μs
Remarks
When the parameter n is set to 1, it indicates External Trigger 1; when set to 2, it indicates External Trigger
2.
This command is only valid when the external trigger delay function is enabled.