-10- 

Using Notices 

1.

 

To use the tracking generator output terminal:

 

 

 

CAUTION

 

To avoid damage to the tracking 

generator, the reverse power cannot 
10 dBm when the frequency 
less than 10 MHz; the reverse power 

cannot 20 dBm when the 
frequency more than 10 MHz. The 

reverse DC voltage cannot exceed 50 V. 

 

2.

 

To use the RF input terminal:

 

 

 

CAUTION

 

To avoid damage to the instrument, 
for the signal input from the RF input 
terminal, the DC voltage component 

and the maximum continuous power 
of the AC (RF) signal component 

cannot exceed 50 V and +20 dBm 
respectively. 

 

Remote Control Overview 

DSA800 supports communication with PC via USB, LAN or GPIB 
(option) interface for remote control. The remote control is 

realized on the basis of SCPI (Standard Commands for 
Programmable Instruments) command set through two ways: 

user-defined programming and PC software (such as 

RIGOL

 

Ultra Sigma). When the instrument is in remote mode, the 

  icon is displayed in the user interface and the front 

panel keys (except 

Esc

) are locked. At this point, you can press 

Esc

 to exit remote mode. 

Summary of Contents for DSA800 Series

Page 1: ...RIGOL 快速指南 Quick Guide 2014 RIGOL TECHNOLOGIES INC All Rights Reserved DSA800 系列频谱分析仪 DSA800 Series Spectrum Analyzer DSA800 ...

Page 2: ...not operate without covers 4 Use proper fuse 5 Avoid circuit or wire exposure 6 Do not operate with suspected failures 7 Keep proper ventilation 8 Do not operate in wet conditions 9 Do not operate in flammable and explosive environment 10 Keep product surface clean and dry 11 Protect the instrument from static electricity 12 Pay attention to handling safety ...

Page 3: ...resolution color LCD display and various remote communication interfaces they can be widely used in various fields such as education company research and development as well as industrial manufacture Preset LCD Menu Softkeys Menu Control Keys Function Keys Knob Print Help Power Switch USB Host Earphone Jack TG Output RF Input Direction Keys Numeric Keyboard AC Connector Fuse Seat Handle Security L...

Page 4: ...20 mm 250V AC T2A fuse CAUTION Make sure that the instrument is properly grounded to avoid electric shock Start up and Self calibration After connecting the instrument to power source correctly press at the front panel to start the spectrum analyzer You can obtain information about the start up initialization process through the start up progress indications Following the start up screen the sweep...

Page 5: ...al the DC voltage component and the maximum continuous power of the AC RF signal component cannot exceed 50 V and 20 dBm respectively Remote Control Overview DSA800 supports communication with PC via USB LAN or GPIB option interface for remote control The remote control is realized on the basis of SCPI Standard Commands for Programmable Instruments command set through two ways user defined program...

Page 6: ...DSA800 Series User s Guide provide detailed introductions of the functions of this product DSA800 Series Programming Guide provide detailed introductions of the SCPI commands and programming of this product DSA800 Series Datasheet provide the main characteristics and specifications of this product DSA800 Series Accessories and Options provide detailed introductions of the accessories and options o...

Reviews: