RIGOL
User’s Guide for DSA700 Series
5-2
Technical Specifications
Frequency
Frequency
DSA705
DSA710
Frequency range
100 kHz to 500 MHz
100 kHz to 1 GHz
Frequency resolution
1 Hz
Internal Reference Frequency
DSA705
DSA710
Reference frequency
10 MHz
Accuracy
±[ (time since last calibration × aging rate) + temperature stability
+ calibration accuracy]
Initial calibration accuracy <1 ppm
Temperature stability
0
℃
to 50
℃
, reference to 25
℃
<2 ppm
Aging rate
<2 ppm/year
Frequency Readout Accuracy
Marker resolution
span/ (number of sweep points - 1)
Marker uncertainty
±(frequency indication × reference frequency ac 1% ×
span + 10% × resolution ban marker resolution)
Frequency Counter
Resolution
1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz
Uncertainty
±(frequency indication × reference frequency accuracy + counter
resolution)
Frequency Span
Range
0 Hz, 100 Hz to maximum frequency of instrument
Uncertainty
±span/ (number of sweep points - 1)