RIGOL
Chapter 2 Command Systems
Programming Guide for DSA1000A series
2-36
Default
10 dB
Front Panel
AMPT, Scale/Div
14.
:DISPlay:WINdow:TRACe:Y:SCALe:RLEVel
Syntax
:DISPlay:WINdow:TRACe:Y:SCALe:RLEVel <ampl>
:DISPlay:WINdow:TRACe:Y:SCALe:RLEVel?
Function
Sets the reference level, such as
:DISP:WIN:TRAC:Y:SCAL:RLEV -10
The query returns “-1.01”.
Explanations
<ampl> ranges from -100 dBm to 30 dBm.
The command may have an influence on the attenuator.
Default
0 dBm
Front Panel
AMPT, Ref Level
15.
:DISPlay:WINdow:TRACe:Y:SCALe:RLEVel:OFFSet
Syntax
:DISPlay:WINdow:TRACe:Y:SCALe:RLEVel:OFFSet <rel_ampl>
:DISPlay:WINdow:TRACe:Y:SCALe:RLEVel:OFFSet?
Function
Sets the reference level offset, such as
:DISP:WIN:TRAC:Y:SCAL:RLEV:OFFS 10
The query returns “1.01”.
Explanations -300 dB to 300 dB
Default
0 dB
Front Panel
AMPT, Ref Offset
16.
:DISPlay:WINdow:TRACe:Y:SCALe:SPACing
Syntax
:DISPlay:WINdow:TRACe:Y:SCALe:SPACing LINear|LOGarithmic
:DISPlay:WINdow:TRACe:Y:SCALe:SPACing?
Function
Sets the scale type.
The query returns “LIN” or “LOG”.
Default
LOGarithmic
Front Panel
AMPT, Scale Type, Log/Lin
17.
:DISPlay:WINdow:TRACe:Y:SCALe:NRLevel
Syntax
:DISPlay:WINdow:TRACe:Y:SCALe:NRLevel <rel_ampl>
:DISPlay:WINdow:TRACe:Y:SCALe:NRLevel?
Function
Sets the reference level of the normalization, such as
:DISP:WIN:TRAC:Y:SCAL:NRL -20
Summary of Contents for DSA1000A Series
Page 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Page 2: ......
Page 14: ......
Page 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Page 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Page 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...