Chapter 2 Command Systems
Programming Guide for DSA1000A series
2-90
Function
Sets the value of X dB to calculate the emission bandwidth, such as
:EBWidth:XDB -20
The query returns “-2.01”.
Explanations
<real_amp> ranges from -100 dB to -0.1 dB.
Default
-10 dB
Front Panel
Meas, EBW
Meas Setup, X dB
67.
[:SENSe]:HDISt:AVERage:COUNt
Syntax
[:SENSe]:HDISt:AVERage:COUNt <integer>
[:SENSe]:HDISt:AVERage:COUNt?
Function
Sets the number of averages of the Harmonic Distortion measurement
results, such as
:HDIS:AVER:COUN 100
The query returns “100”.
Explanations
<integer> ranges from 1 to 1000.
Default
10
Front Panel
Meas, Harmo Dist
Meas Setup, Avg Num
68.
[:SENSe]:HDISt:AVERage[:STATe]
Syntax
[:SENSe]:HDISt:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:HDISt:AVERage[:STATe]?
Function
Turns on or off the average function of the Harmonic Distortion
measurement.
The query returns “0” or “1”.
Default
OFF|0
Front Panel
Meas, Harmo Dist
Meas Setup, Avg Num, On/Off
69.
[:SENSe]:HDISt:AVERage:TCONtrol
Syntax
[:SENSe]:HDISt:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:HDISt:AVERage:TCONtrol?
Function
Sets the average mode of the Harmonic Distortion measurement.
The query returns “EXP” or “REP”.
Explanations
In Exp mode, the result is from the exponent average calculation
of the current N (specified by [:SENSe]:HDISt:AVERage:COUNt)
Summary of Contents for DSA1000A Series
Page 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Page 2: ......
Page 14: ......
Page 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Page 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Page 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...