TM
-
Rightest
Blood Glucose Monitoring System GM700SB is not intended for serum or plasma
testing.
- Inaccurate test results may be obtained at altitudes greater than 10,000 feet (3048 meters) above
sea level.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- The glucose test may be invalid in the presence of abnormally high concentration:
Ascorbic acid
≧
5 mg/dL (0.28 mmol/L), Xylose
≧
20 mg/dL (1.33 mmol/L) and Uric Acid
≧
20
mg/dL (1.19 mmol/L)
.
- Not for screening or diagnosis of diabetes mellitus.
- Not for use on critically ill patients in shock, dehydrated patients or hypo-osmolar patients.
- Alternative site testing (AST) should only be performed during steady-state times (when glucose is
not changing rapidly).
- Alternative site testing should not be used to calibrate continuous glucose monitoring systems
(CGMS).
- Alternative site testing should not be performed if testing for hypoglycemia (Low blood glucose).
- Results from alternative site testing should not be used in insulin dose calculation.
- Regarding potential electromagnetic or other interference, keep meter out of electromagnetic
radiation sources.