![Rigel SafeTest 60 Instruction Manual Download Page 6](http://html1.mh-extra.com/html/rigel/safetest-60/safetest-60_instruction-manual_1460610006.webp)
Rigel SafeTest 60 Instruction Manual
–
Revision 1.2
5
User Notes
These operating instructions are intended for the use of adequately trained
personnel.
Environmental Conditions
The SafeTest 60 has been designed to be operated in a dry environment, at a
temperature of 0 to 40 degrees C without moisture condensation, and at an operating
altitude 0
– 2000m.
The SafeTest 60 has a protection rating of IP40 and is rated for operation at pollution
degree 2 according to IEC 60529.
The following symbols are used in these operating instructions and on the
Rigel SafeTest 60.
Safety Notes
If the SafeTest 60 is used in a manner not specified by these operating instructions then
the protection provided may be impaired.
Only accessories recommended or approved by the manufacturer should be used with
the SafeTest 60.
Do not connect the SafeTest 60 to electrical circuits with nominal voltage greater than
CAT II 300 V AC/DC.
Do not touch test probes beyond the hand barrier on the test probe.
The SafeTest 60 may apply high voltage or mains power to the appliance under test.
Do not touch conductive parts of the appliance while tests are active.
Do not open the SafeTest 60, no user serviceable parts.
Do not operate the SafeTest 60 in an explosive gas or dust environment.
The SafeTest 60 and all associated cables and leads must be checked for signs of
damage before equipment is operated. Do not use if there are signs of damage.
Where safe operation of the SafeTest 60 is no longer possible it should be immediately
shut down and secured to prevent accidental operation.
It must be assumed that safe operation is no longer possible:
- if the instrument or leads show visible signs of damage or
- the instrument does not function or
- after long periods of storage under adverse environmental conditions.
To verify the correct operation of the unit, perform test functions using a known
appliance or checkbox or return the unit to an approved agent for service.
Symbol used for tips and guidance notes in this manual.