Rigel SafeTest 60 Instruction Manual
–
Revision 1.2
20
To activate single fault conditions, use the
and
buttons on the front
panel.
opens the EARTH fault condition relay whilst the
opens the
NEUTRAL fault condition relay.
The leakage screen will indicate the current state of the selected fault conditions.
IEC 61010 tests
Tests
NC
SFC
Touch Leakage
0.5mA
3.5mA
Touch Voltage
33V
55V