14
Figure 14a:
To replace the scale batteries,
push outward on the small tab at the rear
of the scale housing and lift the battery
cover (a small, flat head screwdriver
inserted into the slot at the back of the
lid may be used to gently pry it up). Insert
fresh AA Alkaline batteries; the cells must
be oriented as shown by the inscription in
each battery receptacle.
Scale Information
• To obtain an accurate weight
measurement, the client must be lifted
clear of all weight-bearing surfaces, and
the feet or legs must not be in contact
with the base frame of the TRAM.
• The scale uses four AA Alkaline batteries.
The battery life is approximately
100 hours of continuous use, or
approximately 500 weight recordings.
• The scale will automatically turn off if it
does not sense a change in weight for
more than 10 minutes.
• The scale is accurate to one percent if
used correctly.
• Rifton recommends that the scale be
calibrated by a qualified technician at
three to five year intervals, depending
on frequency of use. For instructions on
service and calibration, please contact
Rifton customer service.
Figure 14a