4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG
Page 22
6.1.8
Fundamental to the methods with reagent blank…
After selecting a method with
reagent blank (RB), the reagent
blank can be measured, entered or
put on zero.
Press [EDIT] to enter the RB
manually or [MEAS.] to measure it.
When using a programmed method
based on a calculation procedure
with RB, the used RB will be stored
together with the corresponding
method number.
After renewed selection of this
method the last stored reagent
blank is offered as “OLD RB”.
Press [YES] to use the last stored
RB or [NO] to enter or measure it.
6.1.9
Fundamental to ID-NO. and sample num
erator…
All test results are labeled with a numerator.
Additionally, all results can be labeled with a 5-digit ID-NO. When the ID-NO. is not zero it will be
displayed and printed together with the sample result.
When a method is selected you can edit both numerator and ID-NO. of a sample with [MODE] [NUM.]
Press [N / I] to switch between
editing numerator and ID-NO.
6.1.10 Fundamental to storing test resul
ts…
All test results are stored automatically. Up to 1000 results can be stored in memory.
See table 7.2.4.1 for the format of the stored data.
Stored results can be output through serial interface (chapter 8.2.8 - Stored results).
When memory is full oldest test results will be overwritten or you can send all results through the serial
interface and then delete them.
O L D R B 0 . 1 1 0
N O Y E S
R E A G E N T B L A N K
E D I T M E A S .
N E W S A M P L E N O . 1 - 9 9
1 2 3 4 5 6 7 8
E S C 9 0 . + - N / I E